background image

Operation

6

Performance Checks

The instrument will operate in any position, but the specified
accuracies assume that the instrument is face up, on a firm
level surface. This is particularly true for hand cranked units
to obtain a smooth constant crank speed. 

1)

Without the test leads being connected to the instrument,

but with the rotary selector switch set to the 1 kV range
(

MJ160

to 500 V range) press and hold down the 

‘Test’

button, whilst turning the generator handle,(

MJ159

MJ160

and

MJ359

) at >180 rev/min.The meter pointer should

remain over the 

’ (infinity) position on the scale. This

establishes that there is no leakage through the instrument
itself.

2)

Check that the test leads, probes and crocodile clips are

in good order, clean and with no broken or cracked insulation.
Connect two of the test leads to the ‘

+

’ and ‘

-

’ terminals on the

side of the instrument case and ensure that their clips are not
touching anything. 

3)

Press the 

‘Test’

button again and keep it pressed whilst

turning the generator handle (Hand cranked models) at >180
rev/min and observe the meter needle. The needle  should
rest over the ‘

’ (infinity) position on the scale. If it does not,

the test leads may be faulty and should be inspected more
closely for damage. Replace them if necessary with

calibrated leads

available as optional accessories 

4)

Connect the test lead clips together, press the 

‘Test’

button and turn the generator handle (hand cranked models)
again. The meter should read zero. If it indicates infinity or a
high resistance value the leads may be open circuit and
should be inspected further. Replace them if necessary. 
(Shorting the leads together and obtaining a zero reading also 

shows that the instrument is working).

Note:

—To avoid creating leakage paths when insulation

testing, it is advisable not to allow the leads to twist together
nor trail across metalwork etc. more than is really necessary.

Voltage measurement

When not testing (i.e. in standby mode) the instruments act as
a voltmeter (0 to 600 Volts a.c.) Therefore, as soon as the test
leads are connected to the item under test, any a.c. voltage
present will be immediately shown. Thus indication is given
that the item has not been completely de-energized.
The instrument also monitors circuit discharge when the

Test

’ button is released following an insulation test on a

capacitive item, e.g. a long cable. In this case it is important
to realize that the actual voltage (d.c. in nature) is not given,
but the meter does indicate when the voltage has decayed to
zero and therefore when it is safe to remove the test leads.
Note, however, that the instrument does not indicate the
presence of negative d.c.voltage.

Resistance Measurement

With the leads connected to the instrument, and having
completed the Preliminary Checks:

1)

Set the selector switch to the ‘

position. 

2)

Connect the leads across the isolated circuit. 

3)

Press and hold the 

‘Test’

button and turn the generator 

handle (hand cranked models) at >180 rev/min. 

4)

The resistance will be indicated on the ‘

scale.

Note:-

If necessary repeat the continuity test with the leads

reversed. The effects of any stray e.m.fs. in the sample under
test may then be negated by taking the average of the two
readings. 

Содержание MJ159

Страница 1: ...Major MEGGER Insulation Testers MJ159 212159 MJ160 212160 MJ359 212359 MJ459 212459 MJ559 212559 User Guide MEGGER...

Страница 2: ...MJ559 5 Battery replacement MJ459 5 Battery Charging MJ559 5 Performance checks 6 Voltage measurement 6 Resistance Measurement 6 Insulation testing 7 Using the Guard terminal 7 Fuse Checking and Repla...

Страница 3: ...cuit connections must not be touched during a test Circuits must be discharged before disconnecting the test leads Replacement fuses must be of the correct type and rating When making a voltage measur...

Страница 4: ...e dayglow needle indicates the resistance value being measured MJ159 and MJ160 are powered by a low voltage hand cranked brushless a c generator which is connected after rectification to a d c to d c...

Страница 5: ...ow Green Neutral White Blue Line Black Brown If using a fused plug a 13 Amp fuse to BS1362 should be fitted PRELIMINARY CHECKS Battery Condition Check MJ459 MJ559 Before initial use of MJ559 charge th...

Страница 6: ...resistance value the leads may be open circuit and should be inspected further Replace them if necessary Shorting the leads together and obtaining a zero reading also shows that the instrument is work...

Страница 7: ...ings are taken after pressing the Test button there is no other control to operate Capacitive circuits automatically discharge through the tester when the Test button is released The approximate disch...

Страница 8: ...sulation Circuit Fuse Check 1 Connect the test leads together and set the rotary selector switch to a M position 2 Press the Test button and keep it pressed whilst turning the generator handle hand cr...

Страница 9: ...250 V 500 V 1000 V 30 0 maximum 50 V 100 V 40 0 maximum Midscale 4 M 2 M for MJ160 Short Circuit Current 1 9 mA 0 65 mA for MJ160 Test Voltage characteristics Accuracy 1 25 of fsd on a 2 8 in 71 1mm a...

Страница 10: ...uracy 2 5 of full scale Power Supplies MJ159 and MJ160 Low voltage brushless a c generator Cranking speed between 130 rpm and 170 rpm MJ359 Dual operation low voltage brushless a c generator or 120 V...

Страница 11: ...11 4 ins x 1 4 ins 32 mm x 6 mm E M C The instruments meet EN 50081 1 and EN 50082 1 1992 Operating Temp Range 32 to 113 F 0 to 45 C for battery operated models 14 to 122 F 10 to 50 C Humidity Range O...

Страница 12: ...L EXTRA Carrying case 6420 043 U S 217740 Test Record Card Pack of 20 6111 216 Black test lead with large alligator clip 6220 295 Red test lead with large alligator clip 6220 586 Green test lead with...

Страница 13: ...Insulation tests are sometimes used in isolation as absolute measures of the quality of the insulation This is most appropriate when equipment is being installed and checked for compliance with a spe...

Страница 14: ...r the surface of the insulation material As the total current depends upon the time for which the voltage is applied Ohm s Law theoretically applies at infinite time The charging current falls relativ...

Страница 15: ...AVO INTERNATIONAL Archcliffe Road Valley Forge Corporate Centre Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8579 Fax 44 0 1304 207342...

Страница 16: ...s MeterCenter com Website http www MeterCenter com This instrument is manufactured in the United Kingdom The company reserves the right to change the specification or design without prior notice MEGGE...

Отзывы: