To delete all records, select “Select All” then select “Delete Record”. 

 

Deleting D Ω String data 

This is impedance data recorded during a discharge test.

On the BITE5 select the record ICON.

  

  

Select “D Ω String”. 

 

www.megger.com

BITE5

52

Deleting D Ω String data

Содержание BITE5

Страница 1: ...BITE5 Battery tester USER MANUAL...

Страница 2: ...ifications are subject to change without notice WARRANTY Products supplied by Megger are warranted against defects in material and workmanship for a period of 1 years following shipment The warranty i...

Страница 3: ...6 Performing an impedance test on a battery string 18 Measuring and recording solar cell voltages and currents 21 Measuring and battery string voltages and currents 24 Performing a discharge test 27 P...

Страница 4: ...1 Deleting meter impedance data 51 Deleting discharge impedance string data 52 Deleting meter VA data 54 Deleting string VA data 55 Deleting discharge VA string data 56 Deleting a string configuration...

Страница 5: ...is manual is written for technical personnel who are familiar with the various measurements performed by volt meters and current meters and have a general understanding of their use and operation Such...

Страница 6: ...BITE5 Battery Tester 1 Duplex Probes 1 Voltage Leads 1 Charger 1 Micro SD Card 1 Micro SD Card Reader 1 Mini USB Cable 1 Neck Strap 1 Zero Bar 1 Stylus Optional AC DC CT Optional PC Bluetooth Dongle...

Страница 7: ...Optional 11 75 mm Tip Concentric Probes Optional 25 4 mm 1 Tip Concentric Probes www megger com BITE5 7 Items received...

Страница 8: ...and operation of the equipment and the hazards involved Safety Precautions The following safety precautions MUST be taken whenever the instrument is installed Wear safety glasses and insulated gloves...

Страница 9: ...X terminal Operating specifications Operating temperature 0 50 C 32 122 F Storage temperature 20 50 C 4 122 F Charging temperature 10 40 C 50 104 F Altitude Operational 0 2000 m Non operational 0 10 0...

Страница 10: ...1 V Current DC AC Range Resolution Accuracy 4 A DC 0 001 A 0 5 of reading 5 digits CT Tolerance 40 A DC 0 01 A 400 A DC 0 1 A 1000 A DC 1 A 4 A AC 0 001 A 0 75 of reading 10 digits CT Tolerance 40 A...

Страница 11: ...ing records Recorded data charts Instrument configuration Zero adjustment For accurate ohmic measurements it is recommended that a zero adjust is performed when changing probes To perform a zero adjus...

Страница 12: ...on zero bar until adjustment is complete Operation Configuration of BITE5 The BITE5 allows you to customize the unit for your needs You can set the desired language the date and time format the scree...

Страница 13: ...or Fahrenheit Then select the Etc tab This screen allows you to customize the settings of your BITE5 From this screen you can select the following Language Set the instrument language Date format Sele...

Страница 14: ...l bluetooth printer SD card Format the micro SD card NOTE This will cause all data and configurations to be erased Reset settings Resets the instrument settings to default factory settings Configurati...

Страница 15: ...tested Lead acid Ni CD Ni MH Li ion Li poly Cell Cell Name Name of string Model Model number of batteries Capacity Battery capacity in Ah or mAh Ref Baseline reference value Warning Warning upper ohm...

Страница 16: ...formed on individual cells or sequentially on battery strings These measurements can be taken on any individual battery up to 200 V DC Operation Measuring and saving individual battery measurements Co...

Страница 17: ...and alarm limit for the impedance value and a lower limit for the voltage This is an optional step Select OK when done Note this feature can be disabled as well by selecting OFF Start testing by plac...

Страница 18: ...and the BITE5 will automatically save any measurement with a date and time stamp Performing an impedance test on a battery string Connect the impedance leads to the input connector of the BITE5 On th...

Страница 19: ...lect desired string Select New Test to start a new test on the selected string Select Select Test if you wish to continue a test that was already in progress www megger com BITE5 19 Performing an impe...

Страница 20: ...erature to memory automatically The results will be displayed on the screen Move to the next battery in the string and take a measurement The recorded values will be displayed on the screen Continue t...

Страница 21: ...es DC or AC panels and UPS output or input voltages The BITE5 will save values for any voltage up to 1000 V DC and 600 V AC Operation Connect the voltage leads to voltage inputs of the BITE5 If measur...

Страница 22: ...measurement V DC and V ripple V DC and Amps DC V AC and Amps AC If using the CT set the correct range on the BITE5 Take measurement www megger com BITE5 22 Measuring and recording solar cell voltages...

Страница 23: ...to manually save the value with a date and time stamp Automatic saving of values Select Auto Hold and the BITE5 will automatically save any measurement with a date and time stamp www megger com BITE5...

Страница 24: ...ing through the string These values will be saved to the selected string data and will have a date and time stamp Operation Connect the voltage leads to voltage inputs of the BITE5 If measuring curren...

Страница 25: ...Select Select desired battery string then press OK Select desired measurement V DC and V ripple V DC and Amps DC V AC and Amps AC www megger com BITE5 25 Measuring and battery string voltages and cur...

Страница 26: ...ct range on the BITE5 Take measurement Press Hold to freeze the measurement on the screen Press Save to manually save the value with a date and time stamp www megger com BITE5 26 Measuring and battery...

Страница 27: ...string and start the discharge test The BITE5 can then be used to take manual measurements of the cell voltage throughout the discharge process In this mode the unit will record the DC voltage of eac...

Страница 28: ...On the BITE5 select VA Select Discharge Select Select Select desired battery string then press OK www megger com BITE5 28 Performing a discharge test...

Страница 29: ...rect range on the BITE5 Take measurement of the first cell The DC voltage and DC current will be saved with a date and time stamp Take measurement of the each following cell Each measurement shall be...

Страница 30: ...es It may not correlate with other causes of cell aging such as plate corrosion In this mode the BITE5 will also measure the cell temperature during the discharge The temperature will be taken off the...

Страница 31: ...Discharge Select SELECT Select desired battery string then either press Select Test to continue a test or select New Test to start a new test www megger com BITE5 31 Performing an impedance and disch...

Страница 32: ...h following cell Each measurement shall be saved in sequence with a cell number date and time stamp When the end of the string is reached the unit will prompt the user to either end the test or select...

Страница 33: ...test Cell voltage trending Trends every voltage value of an individual cell String voltage trending Trends the voltage of all cells in a string for a given test Cell temperature trending Trends every...

Страница 34: ...Select Select Select string then press OK Select desired cell in the left column Select Chart to change the parameter being trended www megger com BITE5 34 Trending recorded impedance data...

Страница 35: ...Trending string data On the BITE5 select the chart ICON Select String Select Select www megger com BITE5 35 Trending recorded impedance data...

Страница 36: ...rge VA cell voltage trending Trends every impedance value of an individual cell Discharge VA string voltage trending Trends the impedance of all cells in a string for a given test Discharge VA cell cu...

Страница 37: ...Operation Trending individual cell data On the BITE5 select the chart ICON Select D VA Cell Select Select www megger com BITE5 37 Trending recorded VA discharge data...

Страница 38: ...n press OK Select desired test in the left column Select Chart to change the parameter being trended Trending string data On the BITE5 select the chart ICON www megger com BITE5 38 Trending recorded V...

Страница 39: ...Select D VA String Select Select Select string then press OK Select the desired test to trend in the left column www megger com BITE5 39 Trending recorded VA discharge data...

Страница 40: ...est Discharge cell impedance trending Trends every voltage value of an individual cell Discharge string impedance trending Trends the voltage of all cells in a string for a given test Discharge cell t...

Страница 41: ...Select D Cell Select Select Select string then press OK Select desired cell in the left column www megger com BITE5 41 Trending recorded impedance discharge data...

Страница 42: ...Select Chart to change the parameter being trended Trending string data On the BITE5 select the chart ICON Select D String www megger com BITE5 42 Trending recorded impedance discharge data...

Страница 43: ...Select Select Select string then press OK Select desired test in the left column Select Chart to change the parameter being trended www megger com BITE5 43 Trending recorded impedance discharge data...

Страница 44: ...orded values of individual impedance measurements made during a discharge test on a string Meter VA These will be the individual recorded voltage and current measurements that were made with the BITE5...

Страница 45: ...s shall be displayed with a date and time stamp Viewing String records On the BITE5 select the record ICON Select String Select desired string then press Select www megger com BITE5 45 Viewing string...

Страница 46: ...sired test in the left column Recorded values will be displayed in the right column Viewing D String records On the BITE5 select the record ICON Select D String www megger com BITE5 46 Viewing D Strin...

Страница 47: ...then press Select Select desired test in the left column Recorded values will be displayed in the right column Viewing Meter VA records On the BITE5 select the record ICON www megger com BITE5 47 Vie...

Страница 48: ...t Meter VA All recorded values will be displayed with a date and time stamp Viewing string VA records On the BITE5 select the record ICON Select String VA www megger com BITE5 48 Viewing string VA rec...

Страница 49: ...then press on Select All recorded values will be displayed with a date and time stamp Viewing D VA String records On the BITE5 select the record ICON Select D VA String www megger com BITE5 49 Viewin...

Страница 50: ...Select desired string then press on Select Select desired test in the left column Recorded values will be displayed in the right column www megger com BITE5 50 Viewing D VA String records...

Страница 51: ...er data This refers to the impedance measurements not associated with a battery string On the BITE5 select the record ICON Select Meter Select desired measurement then select Delete Record www megger...

Страница 52: ...select Select All then select Delete Record Deleting D String data This is impedance data recorded during a discharge test On the BITE5 select the record ICON Select D String www megger com BITE5 52 D...

Страница 53: ...ct Select desired test in the left column then press Delete Test to delete the test To delete an individual record select the desired record on in the right column then select Delete Record www megger...

Страница 54: ...surements not associated with a battery string On the BITE5 select the record ICON Select Meter VA Select desired measurement then select Delete Record To delete all records select Select All then sel...

Страница 55: ...efers to the voltage and current measurements associated with a particular battery string On the BITE5 select the record ICON Select String VA Select desired string then press Select www megger com BI...

Страница 56: ...hen select Delete Record Deleting D VA String data This refers to the recorded voltages taken during a discharge test On the BITE5 select the record ICON Select D VA String www megger com BITE5 56 Del...

Страница 57: ...ess Select Select desired test in the left column then select Delete Test To delete an individual record select the desired record on in the right column then select Delete Record www megger com BITE5...

Страница 58: ...Deleting a string configuration On the BITE5 select the record ICON Select String Select desired string then press Delete String www megger com BITE5 58 Deleting a string configuration...

Страница 59: ...es as bitmaps To do this momentarily press and release the Power ON OFF button The displayed screen shall be saved to the SD card as a bitmap file The bitmap will be located at the following path MEGG...

Страница 60: ...ntric Probes These allow measurements of cells through battery caps that have probe access ports These probes come with either 11 75 mm tips or 25 4 mm 1 tips 90037 562 11 75 mm 1 4 tips 90037 565 25...

Страница 61: ...er adapter Battery charging starts once the power adapter is connected and plugged into AC The battery charge will take approx 4 hours to complete If the unit is operated off of the AC adapter then th...

Страница 62: ...storing for long periods of time there is no need to remove the battery pack However all batteries experience self discharge This will lead to a gradually draining of the batteries For best battery li...

Страница 63: ...rves the right to change the specification or design without prior notice Megger is a registered trademark The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and is...

Отзывы: