Short Cct. Current:

2.75 mA nominal

Temp. Coefficient:

±0.06% per °F

Default Voltage measurement

Range:

0 - 600 V a.c. / d.c. (effective on Insulation ranges 
with ‘Test’ push not pressed)

Input impedance:

300 k

to 1 M

depending on input voltage

Discharge Rate:

A 1 µF capacitor will discharge to a safe level in 

2 s

Accuracy:

±2% of reading ± 1 digit d.c. or a.c. at 45 Hz to 65 Hz

Temp. Coefficient:

±0.06% per °F

General Specifications

Overload rating:

The 

210600

is protected for connection to Power 

distribution systems up to 300 V Line Ground, and 
500 V Line Line for Installation Category 

III

*

.

Temp. Range: 

Operating: 

23°F to 122°F

Storage:

4°F to 158°F 

Humidity:  

Operating:       70% RH at 68°F, 60% RH at 95°F, 50% RH at 104°F 

for measurements <200 M

Storage:

95% RH at 95°F

Display:

3

1

2

digit L.C.D.  Maximum reading 1999

Automatic Discharge:  Capacitive circuits are automatically discharged when 

the ‘Test‘ push is released following an insulation test

Power Supply:

6 x 1.5 V AA ( Alkaline) type cells

Battery life:

Typically 2100 x 5 second operations (worst case)

Note:- Battery cells should not be left in an instrument which may remain
unused for extended periods of time.

Low Battery Indicator:  The low battery indicator symbol

will appear 

when the battery cells are exhausted.

Battery Replacement: The rear cover must not be opened if the test leads
are connected. To remove the rear cover, release the screw at the bottom of
the cover and lift the cover upwards.To avoid the possibility of shock, do not
press the ‘Test‘ push when changing batteries.

Fuses:

2 x 500 mA 500 V (F) H.B.C.10 kA min 1

1

4

“ x 

1

4

To check the +ve fuse, short the +ve and -ve test leads and press the ‘Test
push.  An ‘ ’ reading indicates that the +ve fuse has ruptured.

Fuse Replacement: 

Held in screw type holders located in the base of the

instrument. Use a flat blade screwdriver to release the center part of each
fuse holder. Remove the fuses and test each in turn. Replace with fuse(s) of
the correct type, size and rating. 

Weight:

2.2 lb (including batteries)

Dimensions:

7” x 5” x 5” 

Cleaning:

Wipe disconnected instrument with a clean cloth dampened with 
soapy water or Isopropyl Alcohol (IPA)

*

Relates to  transient overvoltage likely to be found in fixed installation wiring.

0.5A(F) 500V

HBC(10 kA)

Guard fuse

Positive fuse