Tests: Long leads mode
www.megger.com
DLRO2 and DLRO2X
34
8. Tests: Long leads mode
Long leads resistance test mode has been designed for use with test leads of very long length.
The Difference Meter is disabled in long leads mode.
All readings are displayed in mΩ.
Minimum current is 1 A.
Restricted to single tests.
In long leads mode, all soft keys are inactive. The test is pre-set to single measurement,
1 A with previous 2 results displayed.
8.1 Long leads mode - auto and manual start
1. Turn rotary switch to select the long leads mode, either
MANUAL or AUTO position.
NOTE : To AUTO SAVE results click the
SAVE ( ) button
Refer to 5. Data storage (DLRO2X only) on page 22).
OR
Long leads test screen appears, either ‘Long Leads Mode -
Manual Start’ or ‘Long Leads Mode - Auto Start’ will show in
the secondary field for a short time, the text then disappears to
reveal the previous results fields.
2. Connect C1-C2 and P1-P2 to the instrument and the unit
under test.