Raise the unit to its fullest extension.  In most cases, the vanity tray can be removed by pulling 
the tab past the stop.  If the stop is too tight, a slotted screw driver may be used to lightly pry 
the stop.  The vanity can be replaced the same way it was removed. 

 

 

 

 

5.2         Replacing the vanity tray 

   

 

 

 

If replacing the vanity, the rails may need adjustment.  Remove the old vanity as described in 
5.1.  Insert the new vanity. 

If the vanity works well and does not slide too hard or easy, no adjustment is necessary.  If the 
vanity slides too hard or too easily adjustment may be required.  If the vanity needs adjustment, 

start by loosening (3) screws on the outer rail 
with the vanity centered.  If the vanity moved too easy, hold the rail slightly closer to the vanity 
and retighten the screws.  Repeat the process with the other (3) screws on the same rail.  If the 
vanity is too tight, use the same process but move the rail slightly farther away from the vanity.  

 

 

 

Содержание 39" Overbed Table

Страница 1: ...Service Manual 32 Medviron Overbed Table...

Страница 2: ...Table of Contents Section 1 Caster Replacement Section 2 Plug Replacement Section 3 Top Replacement Section 4 Gas Spring Replacement Section 5 Vanity Service...

Страница 3: ...traight slot screw driver 1 0 Turn the over bed table upside down To replace casters that are worn or malfunctioning first raise the unit to its fullest extension Turn the over bed table upside down a...

Страница 4: ...e upside down or on its side To replace damaged or worn end plugs first raise the unit to its fullest extension Turn the over bed table upside down or on its side and place on a protected surface such...

Страница 5: ...esult in voiding of the warranty Tools required 5 32 Allen wrench 2 Square driver bit Provided 3 0 Remove the top Raise the unit to its fullest extension Remove 8 screws from the underside of the plat...

Страница 6: ...t This gas spring keeps the table from being lowered until the handle is activated but allows free rise Gas springs can lose force over time which will cause the table to descend faster when activated...

Страница 7: ...gas spring to the top washer Remove the top nut the top washer set aside Figure 5 Figure 6 4 4 Remove the base Place unit on a soft surface such as a blanket or carpet Remove four bolts washers Anti...

Страница 8: ...arellel to the top washer Figure 5 4 10 Attach the upper column to the top washer Use the provided block to boost the upper column Figure 8 Activate the gas spring lever and push the top plate down al...

Страница 9: ...s described in 5 1 Insert the new vanity If the vanity works well and does not slide too hard or easy no adjustment is necessary If the vanity slides too hard or too easily adjustment may be required...

Отзывы: