1. Trend Arrow
2. The last senor reading or status
3. Time of the last Sensor reading
4. The measuring scope
5. High Limits of glucose reading
6. Low Limits of glucose reading
7. CGM Curve
8. Last glucose point
9. The time scope
Home Screen
Special Conditions
•
Warm-Up
- The Sensor is warming up.
•
ERR
- The Sensor shall be recalibrated after 15 minutes.
•
BG
- The Sensor shall be calibrated now.
•
???
- No readings. Do not remove the Sensor until you get
“Sensor Failure”
Alert.
•
LOST
- Sensor signal has been lost for more than 10 minutes.
Wait for a while.
•
HIGH
- Sensor glucose is above 22.2 mmol/L (400mg/dL).
•
LOW
- Sensor glucose is below 2.2 mmol/L (40mg/dL).
•
Underlined reading -
Calibration overdue. A new meter BG
is needed for calibration .
Read CGM