4.3.3 Live Rhythm Monitor window
The Live Rhythm Monitor window displays ECG, Leadless ECG, Marker Channel, and
telemetered EGM waveform traces. In addition to waveform traces, the Live Rhythm Monitor
shows the following information:
●
The heart rate and the rate interval are displayed if telemetry has been established with
the device.
●
The annotations above the waveform trace show the point at which parameters are
programmed.
The Live Rhythm Monitor appears in the partial view by default, as shown in Figure 9. You
can expand this window to its full size by selecting the small square button in the upper-right
corner of the window or by selecting [Adjust…]. For more information about the Live Rhythm
Monitor, see Section 4.11, “Working with the Live Rhythm Monitor”, page 84.
Figure 9. Live Rhythm Monitor window
1 The location of the square button
2 The location of the [Adjust…] button
4.3.4 Task area
The portion of the screen between the Live Rhythm Monitor window near the top of the
screen and the command bar at the bottom of the screen changes according to the task or
function you select.
One example of a task area is the Parameters screen, which is used to view and program
device parameters as described in Section 4.7, “Viewing and programming device
parameters”, page 70.
Task areas display differently when you perform other functions such as diagnostics and
system tests.