Table 68. Nonprogrammable parameters (continued)
Parameter
Value
Ventricular blanking after high-voltage therapy
520 ms
Fixed bradycardia pacing parameters
Ventricular Safety Pacing intervals
110 ms
PVARP value applied by PVC Response and PMT Intervention
400 ms
NCAP value applied by PVC Response and PMT Intervention
400 ms
Fixed high-voltage therapy parameters
Maximum charging period
30 s
Biphasic
Tilt
50%
Refractory period after ventricular sensed event during cardiover-
sion synchronization
200 ms
Refractory period after paced event during charging or synchroni-
zation
400 ms
Refractory period after charge begins
400 ms
Atrial Vulnerable Period
250 ms
Escape interval after high-voltage therapy
1200 ms
Suspension of VT detection after defibrillation therapy
17 ventricular events
Fixed automatic atrial ATP therapy parameters
VVI/VOO Backup Pacing amplitude
6 V
VVI/VOO Backup Pacing pulse width
1.5 ms
Fixed EP study parameters
T-Shock pacing amplitude
8 V
T-Shock pacing pulse width
1.5 ms
50 Hz burst pacing interval
20 ms
Hardware parameters
Pacing rate limit
(protective feature)
171 bpm
Input impedance
150 kΩ minimum
Recommended Replacement Time (RRT)
Battery Voltage Threshold
≤ 2.63 V
a
35 ms when the ventricular pacing amplitude is programmed to 8 V.
b
The VSP interval may be shortened from 110 ms to 70 ms automatically by the device at higher pacing rates
when necessary to help support ventricular tachycardia detection.
c
PVARP is extended to 400 ms only if the current PVARP is less than 400 ms.
d
The NCAP extension applies only if NCAP is enabled.
e
The waveform for a T-Shock induction can be programmed to biphasic or monophasic.
f
Does not affect event classification during charging.
g
Does not apply during ATP therapies or ventricular safety pacing.
h
If the Upper Tracking Rate or Upper Sensor Rate is programmed to a value greater than 150 bpm, the pacing
rate limit is 200 bpm.
Medtronic
PROTECTA™ XT DR D314DRG
480
Clinician Manual