Table 61. Manual Defibrillation parameters
Parameter
Selectable values
Energy
0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22; 24; 25; 26; 28;
30; 32; 35 J
AX>B; B>AX
a
If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-
voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode is
not used as part of the high-voltage delivery pathway.
Table 62. Manual Cardioversion parameters
Parameter
Selectable values
Chamber
Atrium; RV
Energy
0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22; 24; 25; 26; 28;
30; 32; 35 J
AX>B; B>AX
Minimum R-R (atrial CV only)
400; 410 … 500 … 600 ms
a
If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-
voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode is
not used as part of the high-voltage delivery pathway.
Table 63. Shared manual ATP therapy parameters
Parameter
Selectable values
Minimum Interval (atrial ATP)
100; 110; 120; 130 … 400 ms
Minimum Interval (ventricular ATP)
150; 160 … 200 … 400 ms
Amplitude
1; 2 … 6 ; 8 V
Pulse Width
0.10; 0.20 … 1.50 ms
VVI Backup (for atrial ATP therapy)
On; Off
Pacing Rate
60; 70 … 120 bpm
V. Amplitude
0.50; 0.75 … 5.00; 5.50; 6.00; 8.00 V
0.10; 0.20 … 1.50 ms
a
The default value for this parameter is set according to the permanently programmed settings for bradycardia
pacing.
b
Crosstalk may occur when atrial pacing amplitude is greater than 6.0 V.
Table 64. Manual Ramp therapy parameters
Parameter
Selectable values
Chamber
Atrium; RV
RV Ramp therapy parameters
# Pulses
1; 2 … 6 … 15
%RR Interval
50; 53; 56; 59; 63; 66 … 84; 88; 91; 94; 97 %
Dec/Pulse
0; 10 ; 20; 30; 40 ms
Medtronic
PROTECTA™ XT DR D314DRG
478
Clinician Manual