Table 47. Post Shock Pacing parameters
Parameter
Programmable values
Shipped
Reset
Post Shock A. Amplitude
1; 2; 3; 4 ; 5; 6; 8 V
4 V
4 V
Post Shock A. Pulse
Width
0.1; 0.2 … 1.5 ms
1.5 ms
1.5 ms
Post Shock V. Amplitude
1; 2 … 6 ; 8 V
6 V
6 V
Post Shock V. Pulse
Width
0.1; 0.2 … 1.5 ms
1.5 ms
1.5 ms
a
When tested per CENELEC standard EN 45502-2-1:2003, the tolerance (+40%/-30% for voltages less than 2.0,
and ±30% for voltages greater than or equal to 2.0) is applied not to the programmed setting, but to the calculated
amplitude A, which depends on the programmed amplitude A
p
and programmed pulse width W
p
: A = A
p
x [0.9 –
(W
p
x 0.145 ms
-1
)]
b
When tested per CENELEC standard EN 45502-2-1:2003, the measured pulse width W depends on the load
Rload (in Ohms) and programmed pulse width W
p
(in seconds): W ≤ W
p
+ 34 µs and W ≥ the smaller of (W
p
-
16 µs) or (124 µs + (4 µs x Rload)).
Table 48. Rate Drop Response parameters
Parameter
Programmable values
Shipped
Reset
Rate Drop Response
On; Off
Off
Off
Detection Type
Drop ; Low Rate; Both
Drop
Drop
Drop Size
10; 15 … 25 … 50 bpm
25 bpm
25 bpm
Drop Rate
30; 40 … 60 … 100 bpm
60 bpm
60 bpm
Detection Window
10; 15; 20; 25; 30 s
1 ; 1.5; 2; 2.5 min
1 min
1 min
Detection Beats
1; 2; 3 beats
3 beats
3 beats
Intervention Rate
70; 75 … 100 … 150 bpm
100 bpm
100 bpm
Intervention Duration
1; 2 … 15 min
2 min
2 min
a
When Rate Drop Response is set to On, the lower rate is automatically set to 45 bpm.
Table 49. Sleep parameters
Parameter
Programmable values
Shipped
Reset
Sleep
On; Off
Off
Off
Sleep Rate
30; 35 … 50 ; 55; 60; 70; 75 …
100 bpm
50 bpm
50 bpm
Bed Time
00:00; 00:10 … 22:00 … 23:50
22:00
22:00
Wake Time
00:00; 00:10 … 07:00 … 23:50
07:00
07:00
Table 50. Non-Competitive Atrial Pacing (NCAP) parameters
Parameter
Programmable values
Shipped
Reset
Non-Comp Atrial Pacing On ; Off
On
On
NCAP Interval
200; 250; 300 ; 350; 400 ms
300 ms
300 ms
Medtronic
PROTECTA™ XT DR D314DRG
470
Clinician Manual