Table 38. Blanking periods
Parameter
Programmable values
Shipped
Reset
PVAB Interval
10; 20 … 150 … 300 ms
100; 110 … 150 … 300 ms
150 ms
150 ms
PVAB Method
Partial ; ; Absolute
Partial
Partial
A. Blank Post AP
150; 160 … 200 … 250 ms
200 ms
240 ms
A. Blank Post AS
100 ; 110 … 170 ms
100 ms
100 ms
V. Blank Post VP
150; 160 … 200 … 450 ms
200 ms
240 ms
V. Blank Post VS
120 ; 130 … 170 ms
120 ms
120 ms
a
When PVAB Method = or Absolute
b
When PVAB Method = Partial
c
Programming the PVAB method to Absolute will automatically reset the interval to 30 ms. If the PVAB method
is programmed to Partial or the interval will automatically be reset to 150 ms.
Table 39. Rate Response Pacing parameters
Parameter
Programmable values
Shipped
Reset
Upper Sensor Rate
80; 85 … 120 … 175 bpm
120 bpm
120 bpm
ADL Rate
60; 65 … 95 … 170 bpm
95 bpm
95 bpm
Rate Profile Optimization On ; Off
On
Off
ADL Response
1; 2; 3 ; 4; 5
3
3
Exertion Response
1; 2; 3 ; 4; 5
3
3
Activity Threshold
Low; Medium Low ; Medium High;
High
Medium Low
Medium Low
Activity Acceleration
15; 30 ; 60 s
30 s
30 s
Activity Deceleration
Exercise ; 2.5; 5; 10 min
Exercise
5 min
ADL Setpoint
5; 6 … 40; 42 … 80
18
18
UR Setpoint
15; 16 … 40; 42 … 80; 85 … 180
40
40
Table 40. Rate Adaptive AV parameters
Parameter
Programmable values
Shipped
Reset
Rate Adaptive AV
Off ; On
Off
On
Start Rate
50; 55 … 90 … 145 bpm
80 bpm
60 bpm
Stop Rate
55; 60 … 130 … 175 bpm
130 bpm
120 bpm
Minimum Paced AV
30; 40 … 140 … 200 ms
140 ms
140 ms
Minimum Sensed AV
30; 40 … 110 … 200 ms
110 ms
110 ms
Medtronic
PROTECTA™ XT DR D314DRG
468
Clinician Manual