Table 15. Projected service life in years
Percent
pacing
Maximum
energy
charging
Pre-arrhyth-
mia EGM
storage
500 Ω pacing
impedance
600 Ω pacing
impedance
900 Ω pacing
impedance
2.5 V
3.5 V
2.5 V
3.5 V
2.5 V
3.5 V
DDD,
0%
Semi-annual Off
8.0
8.0
8.0
8.0
8.0
8.0
Quarterly
Off
7.1
7.1
7.1
7.1
7.1
7.1
Semi-annual On
7.8
7.8
7.8
7.8
7.8
7.8
Quarterly
On
7.0
7.0
7.0
7.0
7.0
7.0
DDD,
15%
Semi-annual Off
7.7
7.3
7.7
7.4
7.8
7.6
Quarterly
Off
6.9
6.6
6.9
6.7
7.0
6.8
Semi-annual On
7.5
7.2
7.6
7.3
7.6
7.4
Quarterly
On
6.7
6.5
6.8
6.5
6.8
6.7
DDD,
50%
Semi-annual Off
7.0
6.1
7.1
6.4
7.4
6.8
Quarterly
Off
6.3
5.6
6.4
5.8
6.6
6.2
Semi-annual On
6.9
6.0
7.0
6.2
7.2
6.7
Quarterly
On
6.2
5.5
6.3
5.7
6.5
6.0
DDD,
100%
Semi-annual Off
6.2
5.0
6.5
5.3
6.9
5.9
Quarterly
Off
5.7
4.7
5.9
4.9
6.2
5.4
Semi-annual On
6.1
4.9
6.3
5.2
6.7
5.8
Quarterly
On
5.6
4.6
5.8
4.8
6.1
5.3
AAI<=>
DDD
(MVP
mode)
50%
Atrial,
5% Ven-
tricular
Semi-annual Off
7.5
7.0
7.6
7.1
7.8
7.4
Quarterly
Off
6.8
6.3
6.8
6.4
7.0
6.7
Semi-annual On
7.4
6.8
7.5
7.0
7.6
7.3
Quarterly
On
6.6
6.2
6.7
6.3
6.8
6.5
a
Maximum energy charging frequency may include full energy therapy shocks or capacitor formations. Additional
full-energy charges due to therapy shocks, device testing, or capacitor formation reduces longevity by
approximately 32 days (0.09 years).
b
The data provided for programming Pre-arrhythmia EGM storage to On is based on a 6-month period (two
3-month follow-up intervals) over the life of the device. Additional use of Pre-arrhythmia EGM storage reduces
projected service life by approximately 19% or 2.3 months per year.
Note: These projections are based on typical shelf storage time. Assuming worst-case shelf
storage time (18 months), longevity is reduced by approximately 7.1%.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
449