If you perform an atrial Fixed Burst induction, you may choose to have the device deliver
VVI Backup pacing.
10.7.5.1 Considerations for inducing AT or VT with Fixed Burst
Warning: Monitor the patient carefully when using an EP study function. Have an external
defibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmia
may degenerate to ventricular fibrillation.
Atrial Amplitude and VVI Backup pacing – VVI Backup pacing during an atrial Fixed
Burst induction may be inhibited by crosstalk if the test value for atrial Amplitude is greater
than 6 V.
10.7.5.2 How to deliver a Fixed Burst induction
Figure 223. Fixed Burst induction screen
1. Select Tests > EP Study.
2. Select Fixed Burst from the list of inductions and therapies.
3. If the Select Chamber dialog box appears, select [Atrium] or [RV].
4. If you want to treat the induced episode with a manual therapy, select [Suspend] to
prevent automatic detection.
Note: The [Suspend] button is located at the top of the screen and is not shown in
5. Select the Resume at BURST check box for automatic detection and therapy, or clear
the check box for manual therapy.
6. Accept the displayed test values or select new test values.
Medtronic
PROTECTA™ XT DR D314DRG
442
Clinician Manual