8.5.2 Operation of the Onset feature
Figure 168. Gradual versus sudden rate acceleration
Interval (ms)
Time (s)
Time (s)
Interval (ms)
VT detection
VT detection withheld
Sinus tachycardia
Ventricular tachycardia
200
400
600
900
1200
1500
-12
-10
-8
-6
-4
-2
-0
200
400
600
900
1200
1500
-12
-10
-8
-6
-4
-2
-0
A programmable Onset Percent value is used to evaluate the suddenness or gradualness
of the change in average cycle length from one set of 4 intervals to the next. If you program
the Onset Percent value lower, a larger rate acceleration will be required for the device to
detect VT. If you program the Onset Percent value higher, a smaller rate acceleration will
be required for the device to detect VT.
If the ventricular rate accelerates gradually, as shown in Figure 169, then the Onset feature
prevents sensed ventricular intervals that occur in the VT detection zone from being
classified as VT events.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
335