Figure 169. Operation of Onset during a gradually accelerating ventricular rate
A
S
A
S
V
S
V
S
V
S
V
S
V
S
V
S
A
S
A
S
A
S
A
S
A
S
A
S
A
S
A
S
A
S
V
S
V
S
V
S
V
S
V
S
V
S
4
1
0
4
1
0
4
0
0
4
0
0
4
0
0
3
9
0
3
9
0
3
9
0
3
8
0
3
8
0
3
8
0
0
0
0
0
0
0
200 ms
417
432
96%
415
427
97%
412
425
96%
407
422
96%
405
417
97%
402
415
96%
397
412
96%
395
407
97%
392
405
96%
387
402
96%
385
397
96%
382
395
96%
Marker Channel
V-V interval
VT event count
VT interval
Onset calculation
ECG
1 The ventricular rate is accelerating (interval cycle lengths are decreasing).
2 The ventricular rate is now in the VT detection zone, but the acceleration is gradual. The average
cycle length of any set of 4 intervals is never 81% or less of the previous set of 4 intervals. (81%
is the Onset Percent value in this example.)
If the ventricular rate accelerates rapidly, as shown in Figure 170, then the Onset feature
permits sensed ventricular intervals that occur in the VT detection zone to be classified as
VT events.
Medtronic
PROTECTA™ XT DR D314DRG
336
Clinician Manual