Caution: The device provides no pacing support when it is programmed to ODO mode.
Use ODO mode only in clinical situations where bradycardia pacing is not necessary or is
detrimental to the patient.
7.2.3.5 DOO mode
The DOO mode provides AV sequential pacing at the programmed Lower Rate with no
inhibition by intrinsic events.
Warning: The device provides no sensing or detection in either chamber when it is
programmed to DOO mode. Use DOO mode only in situations in which asynchronous
pacing is warranted.
To program the device to the DOO mode, VT Detection and VF Detection must be
programmed to Off, and AT/AF Detection must be programmed to Monitor.
7.2.4 Operation of single chamber pacing
Single chamber pacing modes are used to pace either the atrium or the ventricle.
7.2.4.1 AAIR<=>DDDR and AAI<=>DDD modes
For information about the AAIR<=>DDDR and AAI<=>DDD modes (MVP modes), see
Section 7.3, “Reducing unnecessary ventricular pacing with MVP mode”, page 223.
7.2.4.2 VVIR and VVI modes
In the VVIR and VVI modes, the ventricle is paced if no intrinsic ventricular events are
sensed. Pacing occurs at the programmed Lower Rate in the VVI mode and at the sensor
rate in the VVIR mode (see Figure 72). In VVIR and VVI modes, the device continues sensing
atrial events for tachyarrhythmia detection purposes.
Medtronic
PROTECTA™ XT DR D314DRG
216
Clinician Manual