Figure 72. Operation of single chamber ventricular pacing in VVIR
200 ms
A
S
V
P
V
P
A
S
V
P
V
S
V
P
ECG
Marker Channel
Sensor interval
1 A ventricular paced event occurs when no intrinsic ventricular event is sensed.
7.2.4.3 AAIR and AAI modes
In the AAIR and AAI modes, the atrium is paced if no intrinsic atrial events are sensed.
Pacing occurs at the programmed Lower Rate in the AAI mode and at the sensor rate in the
AAIR mode (see Figure 73).
A sensed event that occurs during the Atrial Refractory Period is classified as refractory and
does not inhibit atrial pacing. In AAIR and AAI modes, the device continues sensing
ventricular events for tachyarrhythmia detection purposes. VT/VF detection is compromised
in the AAIR and AAI modes. Cross-chamber blanking can cause ventricular events to go
undetected, and crosstalk can cause false detection.
Warning: Do not use the AAIR or AAI mode in patients with impaired AV nodal conduction
because these modes do not provide ventricular support.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
217