![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 204](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148204.webp)
Figure 65. Programmable blanking periods
Atrial blanking
Ventricular
blanking
Marker Channel
A
S
A
P
V
P
V
S
1 For the duration of this atrial blanking period, which is defined by the A. Blank Post AS parameter,
atrial sensing is disabled after a sensed atrial event.
2 For the duration of this ventricular blanking period, which is defined by the V. Blank Post VS
parameter, ventricular sensing is disabled after a sensed ventricular event.
3 For the duration of this atrial blanking period, which is defined by the A. Blank Post AP parameter,
atrial sensing is disabled after a paced atrial event.
4 For the duration of this ventricular blanking period, which is defined by the V. Blank Post VP
parameter, ventricular sensing is disabled after a paced ventricular event.
The cross-chamber blanking periods listed in Table 12 are nonprogrammable.
Table 12. Cross-chamber blanking periods
Parameter
Value
Atrial blanking after a ventricular pacing pulse
30 ms
Ventricular blanking after an atrial pacing pulse
30 ms
a
If the RV pacing amplitude is programmed at 8 V, this value is 35 ms.
The post-shock blanking periods are also nonprogrammable. After a cardioversion or
defibrillation therapy is delivered, the atrial and ventricular blanking is 520 ms.
7.1.4 Operation of Post-Ventricular Atrial Blanking (PVAB)
The system uses Post-Ventricular Atrial Blanking (PVAB) to eliminate the effect of far-field
R-waves. Far-field R-waves are ventricular events that are sensed in the atrium. The PVAB
operation is determined by 2 programmable parameters: PVAB Interval and PVAB Method.
Atrial events that are sensed during the PVAB interval are used only by tachyarrhythmia
detection and do not affect pacing timing. However, changing the PVAB interval determines
whether or not events fall in the interval.