The expected service life of the device after RRT, defined as the Prolonged Service Period
(PSP), is 3 months (90 days).
After the 90-day PSP has expired, the device reaches End
of Service (EOS), and the programmer displays the EOS symbol.
Warning: Replace the device immediately if the programmer displays an EOS indicator.
The device may lose the ability to pace, sense, and deliver therapy adequately after the
EOS indicator appears.
6.13.1.3 Capacitor charging and high voltage therapy information
The Battery and Lead Measurement screen reports information about the last high voltage
charge, the last capacitor formation, and the last delivered high voltage therapy. The Last
Charge section displays the date, charge time and energy range from the last time the high
voltage capacitors were charged (from any starting energy to any final energy). The Last
Capacitor Formation section reports the date, charge time and energy range from the last
time the device capacitors were charged to full energy and that charge remained on the
capacitors for at least 10 minutes. The Last High Voltage Therapy section reports the date,
measured impedance, delivered energy, waveform and pathway for the last delivered high
voltage therapy.
6.13.1.4 Sensing Integrity Counter
When the device senses high-frequency electrical noise, the result is often a large number
of ventricular sensed events with intervals near the programmed value for ventricular
blanking after a ventricular sense (V. Blank Post VS). The Sensing Integrity Counter records
the number of ventricular events with intervals that are within 20 ms of the V. Blank Post VS
parameter value. A large number of short ventricular intervals may indicate oversensing,
lead fracture, or a loose setscrew. If the Sensing Integrity Counter reports more than 300
short ventricular intervals, investigate potential sensing and lead integrity issues. For more
information, see Section 6.4, “Monitoring leads using RV Lead Integrity Alert”, page 141,
and Section 8.8, “Discriminating RV lead noise from VT/VF”, page 345.
8
EOS may be indicated before the end of 90 days if the actual battery usage exceeds the expected conditions
during the Prolonged Service Period. For an explanation of these conditions, see Section A.2, “Replacement
9
EOS may also be indicated if an excessive charge time occurs.
Medtronic
PROTECTA™ XT DR D314DRG
188
Clinician Manual