The Battery and Lead Measurements screen and its associated printed report provide the
most recent battery voltage measurement, as well as the RRT indicator with date and time,
if applicable.
Full Charge information shows the date and duration of the last time the capacitors were
charged from 0 J to full energy.
Select the [>>] button to see more detailed battery and lead measurement data. For more
information, including the date when the battery reaches RRT, see Section 6.13, “Viewing
detailed device and lead performance data”, page 186.
Lead status and trends – Information about lead status allows you to assess the
performance and integrity of leads and identify any unusual conditions. The “Last Measured”
column shows the most recently measured lead impedance for each lead.
Select the [>>] button in the “Last Measured” column to see more detailed lead
measurements and relevant programmed settings.
The lead trend graphs on the Quick Look II screen show lead impedance, capture threshold,
and sensing amplitude measurements recorded over the last 12 months.
Select the [>>] button beside any of the lead trend graphs to see more detailed information
about lead performance. The detailed trend graphs display up to 15 of the most recent daily
measurements and up to 80 weekly summary measurements (showing minimum,
maximum, and average values for each week).
For more information about lead performance graphs, see Section 6.13, “Viewing detailed
device and lead performance data”, page 186.
6.2.2.2 Assessing the patient’s condition
Pacing and sensing information – This information can help to assess the patient’s AV
conduction status and evaluate the effectiveness of programmed device settings.
Information about atrial and ventricular pacing and sensing is shown as percentages of the
total time during the reporting period. This includes the percentage of time that AS-VS,
AS-VP, AP-VS, and AP-VP event sequences occurred.
“MVP On” and “MVP Off” refer to the currently programmed pacing mode, not the usage of
MVP mode pacing since the last session. If the device was programmed to an MVP mode
during the reporting period, a high percentage of ventricular pacing may indicate that the
patient has heart block.
The percentage of time that the patient experienced AT/AF can help you to assess the need
to adjust the patient’s device or drug-based therapies. The time in AT/AF is calculated from
the point of AT/AF onset. For more information, see Section 8.1, “Detecting atrial
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
131