13. To review the stored data for the induced episode, select [Retrieve Data…]. To view
more details, print a Last VT/VF with EGM report, or select Data > Clinical
Diagnostics > Arrhythmia Episodes to view the data on the programmer.
14. Select [Adjust Permanent…] to program a new VF Therapy Rx1 energy level or to
change the Pathway, if desired.
15. Wait until the on-screen timer reaches 5 min, then repeat Step 10 through Step 15 as
needed.
16. Before closing the pocket, select the Params icon and program VF Detection, FVT
Detection, and VT Detection to Off.
5.5.4 How to perform defibrillation threshold testing using 50 Hz Burst
1. Select Tests > EP Study.
2. Select 50 Hz Burst from the list of EP Study functions.
3. Select [RV] in the Select Chamber box.
4. Confirm that the Resume at BURST check box is selected to resume arrhythmia
detection after the induction is delivered.
Note: During a wireless telemetry session, you cannot deliver a 50 Hz Burst induction
when there is a magnet or programming head over the device and the Resume at
BURST check box is selected. If an error message appears, remove the magnet or
programming head, or clear the Resume at BURST check box.
5. Select [Adjust Permanent…].
6. Set the Energy parameter for VF Therapy Rx1 to 10 J less than the desired final
programmed value. Set VF Therapies Rx2–Rx6 to the maximum value.
7. Set the RV Sensitivity parameter to a value that results in an adequate safety margin
for detecting VF. For a final programmed RV Sensitivity of 0.3 mV, an adequate safety
margin is typically attained by setting the value to 1.2 mV during testing.
8. Set VF Enable to On. This also automatically sets the AF/Afl, Sinus Tach, and Wavelet
features to On.
9. Select [PROGRAM].
10. Select [Close].
11. Press and hold the [50 Hz BURST Press and Hold] button. Remove the touch pen from
the [50 Hz BURST Press and Hold] button to automatically abort the induction or
therapy.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual