Figure 23. Inserting the torque wrench into the setscrew
1a
1b
2. Push the lead connector or pin plug into the connector port until the lead connector pin
is clearly visible in the pin viewing area. If necessary, sterile water may be used as a
lubricant. No sealant is required.
3. Confirm that the lead is fully inserted into the connector pin cavity by viewing the device
connector block from the side or end.
a. The lead connector pin should be clearly visible beyond the setscrew block (see
b. The lead connector ring should be completely inside the spring contact block.
There is no setscrew in this location (see Figure 24).
Figure 24. Confirming the lead connection
3b
3a
4. Tighten the setscrew by turning it clockwise until the torque wrench clicks. Remove
the torque wrench.
5. Gently tug on the lead to confirm a secure fit. Do not pull on the lead until the setscrew
has been tightened.
6. Repeat these steps for each lead.
Medtronic
PROTECTA™ XT DR D314DRG
116
Clinician Manual