5.2.1 Selecting the leads
Transvenous or epicardial leads may be used. Do not use any lead with this device without
first verifying lead and connector compatibility.
The device is typically implanted with the following leads:
●
1 quadripolar/tripolar transvenous lead with a trifurcated/bifurcated connector in the
right ventricle (RV) for sensing, pacing, and cardioversion/defibrillation therapies
●
1 bipolar transvenous lead in the atrium (A) for sensing and pacing. Use of a bipolar
atrial lead with ring and tip electrodes spaced ≤ 10 mm apart to reduce far-field R-wave
sensing is recommended.
Note: Use of an epicardial defibrillation lead instead of an RVcoil (HVB) lead may adversely
affect the results of the OptiVol fluid measurements.
5.2.2 How to verify lead and connector compatibility
Warning: Verify lead and connector compatibility before using a lead with this device. Using
an incompatible lead may damage the connector, resulting in electrical current leakage or
resulting in an intermittent electrical connection.
Note: Medtronic 3.2 mm low-profile leads are not directly compatible with the device IS-1
connector block.
Note: If you are using a lead that requires an adaptor for this device, please contact your
Medtronic representative for information about compatible lead adaptors.
Use the information in Table 9 to select a compatible lead.
Table 9. Lead and connector compatibility
Connector port
Primary lead
RV (HVB), SVC (HVX)
DF-1
A, RV
bipolar
a
DF-1 refers to the international standard ISO 11318:2002.
b
IS-1 refers to the international standard ISO 5841-3:2000.
5.2.3 Implanting the leads
Implant the leads according to the technical manuals supplied with the leads unless suitable
chronic leads are already in place.
Warning: Pinching the lead can damage the lead conductor or insulation, which may cause
unwanted high-voltage therapies or result in the loss of sensing or pacing therapy.
Medtronic
PROTECTA™ XT DR D314DRG
112
Clinician Manual