5 Implanting the device
5.1 Preparing for an implant
The following implant procedures are provided for reference only. Proper surgical
procedures and sterile techniques are the responsibility of the physician. Each physician
must apply the information in these procedures according to professional medical training
and experience.
For information about replacing a previously implanted device, see Section 5.8, “Replacing
a device”, page 123.
Ensure that you have all of the necessary instruments, system components, and sterile
accessories to perform the implant.
5.1.1 Instruments, components, and accessories required for an implant
The following non-implanted instruments are used to support the implant procedure:
●
Medtronic CareLink Model 2090 Programmer with Conexus telemetry and a
Conexus Activator, or a Medtronic CareLink Model 2090 Programmer with a
Model 2067 or 2067L programming head
●
Programmer software application for the Protecta XT DR Model D314DRG device
●
Model 2290 Analyzer or equivalent pacing system analyzer
●
external defibrillator
The following sterile system components and accessories are used to perform the implant:
●
implantable device and lead system components
●
programming head sleeve (if a programming head is used)
Note: If a sterilized programming head is used during an implant, a sterile programming
head sleeve is not necessary.
●
pacing system analyzer cables
●
lead introducers appropriate for the lead system
●
extra stylets of appropriate length and shape
4
The software application, Model SW009, can be installed by your Medtronic representative.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
109