Battery Support System 2 Operating Instructions
1-3
©1999–2004 Medtronic Emergency Response Systems, Inc.
Safety Information
Note:
The
Battery Support System
2 has no ON/OFF switch. When connecting the system to ac line
voltage, ensure that the power cord is easily accessible and is not obstructed in any way. If the system
malfunctions or overheats, disconnect the system from the ac line voltage by removing the power cord.
SYMBOLS AND ABBREVIATIONS
The symbols and abbreviations below may be found in these operating instructions, on the
Battery
Support System
2, or on FASTPAK and LIFEPAK
®
batteries:
Possible or potential inaccurate battery charge level indicator.
Using the Battery Support System (MIN 801807) or the two-well Battery Charger (MINs 9-00284,
9-00288, and 801530) to charge and maintain a FASTPAK
®
2 battery will eventually result in an
inaccurate battery charge level indicator. Use only the Battery Support System 2 (MIN 3010035) to
charge and maintain FASTPAK 2 batteries.
Possible equipment damage.
The supplied power cord is used to connect the battery support system to ac line voltage. Excessive
voltage may cause fuses in the Battery Support System 2 to blow. Before connecting the battery
support system, ensure that you are using a grounded outlet of the correct voltage.
Attention, consult accompanying documents
Recycle NiCd battery
Recycle PB battery
See instructions for recycling procedures,
4-6
See instructions for disposal procedure,
3-5
Direct Current (DC)
Alternating Current (AC)
Positive terminal
Negative terminal
Fuse
CAUTIONS!
or