AR-I-MREMSW-5 Avant REM Speech Training Manual 

 

   

Effective Date: 4/07/2014 
 

 

12 

Using the MedRx Probe Tube Stabilization System

 

To assist in maintaining proper probe tube position 
during Live Speech Mapping and Real Ear 
Measurement, MedRx has developed the Probe 
Tube Stabilization System.   

After bending the ear hook and attaching the probe 
assembly and tube, slide the Stabilization System 
onto the ear hook as shown.  Align the larger hole 
with the ear canal and thread the probe tube 
through.  The black tube depth marker should rest 
just inside the larger hole. 

 

 

 

Positioning the Speakers

 

Proper positioning of the 
loudspeakers relative to 
the patient’s hearing 
aids is critical for making 
accurate and repeatable 
measurements. Seat the 
patient so their ear is 
between 19 and 39 
inches from the 
loudspeaker.   

 

Make sure the grill of the loudspeaker is level with and directly in front of the patient’s head. 

When using Live Speech inputs, position the patient so the mouth of the person speaking is in the same 
position as the loudspeaker shown here. 

 

Содержание Avant REM Speech

Страница 1: ...TRAINING MANUAL...

Страница 2: ...AR I MREMSW 5 Avant REM Speech Training Manual Effective Date 4 07 2014 2 This Page Intentionally Left Blank...

Страница 3: ...g 9 Live Speech Mapping 13 Real Ear Measurements 18 Additional Tests and Features 20 Printing 26 Limited Warranty 26 0086 www medrx usa com 1200 Starkey Rd 105 Largo FL 33771 U S A Toll Free 888 392 1...

Страница 4: ...peech Mapping MedRx s in situ method for getting the fitting right the first time every time This manual assumes that the system hardware software and drivers are installed and working properly Please...

Страница 5: ...ext sensitive interactive Help System To access this tool press the F1 key at any time or click the Help menu with your mouse This will display additional information and instructions for the specific...

Страница 6: ...Speech main program by clicking on the Options text menu as shown below The software allows you to set several default options as described below The View tab allows you to set the default appearance...

Страница 7: ...ok and behavior of the Master Hearing Aid and Hearing Loss Simulator If your system has a Tympanometer connected this tab will assist in configuring the look of the display NOTE Remember that more inf...

Страница 8: ...tab are described below For more in depth information consult the interactive Help system within the software The General Tab configures the overall Avant REM Speech behavior with respect to other Wi...

Страница 9: ...wn above The cables will only connect when the white mark on the cable connector aligns with the black mark on the main unit DO NOT force the connector on 1 Locate the Probe Microphone case 2 Carefull...

Страница 10: ...ttaching the Probe Microphone Assembly to the Ear Hook 1 Hold the probe microphone assembly so that with the ear hook facing the front of the head the microphone ports face away from the patient 2 Car...

Страница 11: ...Function Calibrate Probe Tube Position probe microphone between 19 and 39 inches from the loudspeaker and click Start The signal must be at least 70 dB to complete a valid calibration If the screen p...

Страница 12: ...the larger hole with the ear canal and thread the probe tube through The black tube depth marker should rest just inside the larger hole Positioning the Speakers Proper positioning of the loudspeakers...

Страница 13: ...ram was entered using another manufacturer s audiogram module If none of these have occurred you may manually enter the audiometric data into the AVANT REM Speech software as follows 1 Launch the audi...

Страница 14: ...meet the needs of your practice Speech Mapping with Recorded Inputs For those who prefer a standard repeatable stimulus for Speech Mapping the software provides several digital audio files These are a...

Страница 15: ...hese features please consult the Interactive Help system by pressing the F1 key or clicking the Help icon 1 Reference Microphone Level 2 Ear selection 3 Measurement type a Peaks b LTAS 4 Input Selecto...

Страница 16: ...by opening the Avant REM Speech software from within NOAH as shown above From the Avant REM Speech main screen click the NOAH LSM button The REM Speech window will reduce in size and float on top of...

Страница 17: ...g values b Pure Tone Air and Bone conduction thresholds c Narrow Band Noise or Pure Tone UCL at 500 2000 and 3000 Hz d Other tests if needed in your practice or jurisdiction 2 Complete a careful otosc...

Страница 18: ...nerate targets The following rules are supported Gain POGO I POGO II Berger 1 3 Gain NAL RP FIG6 DSL NAL NL1 Real Ear Unaided Response Measures the additive effect of the ear canal s resonance to a st...

Страница 19: ...n is useful in determining the acoustic seal of a hearing aid relative to a sound field outside the ear Some clients report they seem like they are speaking in a barrel or feel blocked up These are si...

Страница 20: ...the hearing instrument If RECD data is present the software will display an option to choose between Ear or Coupler measurement on all Real Ear screens where measurements in a coupler are appropriate...

Страница 21: ...aring instrument is used be sure to perform this measurement at the same volume control setting as the first measurement Select START to open the measurement type selection screen Choose the Coupler R...

Страница 22: ...ve at 1 24 octave Using this information make any needed adjustments to the hearing instrument to reduce or remove the feedback Hearing Loss Simulator The Hearing Loss Simulator demonstrates the effec...

Страница 23: ...al source and the task icons for Normal and Simulate The Master Hearing Aid applies gain separately for each ear If audiogram values are entered for the left and right ear the selected fitting rules w...

Страница 24: ...odify the input signal according to the selected rule This selection allows the patient to hear the benefits of amplification Toggle between these two selections as necessary 6 If the simulated freque...

Страница 25: ...nu click one of the following options Print Setup selects the printer you want the report to be sent to Avant REM Speech prints to any installed Windows printer Page Setup Allows you to set margin siz...

Страница 26: ...as provided under this warranty is the sole and exclusive remedy of the purchaser MedRx shall not be liable for any consequential or incidental damages or for breach of any express or implied warrant...

Отзывы: