AR-I-MREMSW-5 Avant REM Speech Training Manual
Effective Date: 4/07/2014
12
Using the MedRx Probe Tube Stabilization System
To assist in maintaining proper probe tube position
during Live Speech Mapping and Real Ear
Measurement, MedRx has developed the Probe
Tube Stabilization System.
After bending the ear hook and attaching the probe
assembly and tube, slide the Stabilization System
onto the ear hook as shown. Align the larger hole
with the ear canal and thread the probe tube
through. The black tube depth marker should rest
just inside the larger hole.
Positioning the Speakers
Proper positioning of the
loudspeakers relative to
the patient’s hearing
aids is critical for making
accurate and repeatable
measurements. Seat the
patient so their ear is
between 19 and 39
inches from the
loudspeaker.
Make sure the grill of the loudspeaker is level with and directly in front of the patient’s head.
When using Live Speech inputs, position the patient so the mouth of the person speaking is in the same
position as the loudspeaker shown here.