Medrad Stellant D Скачать руководство пользователя страница 6

LSUHSC Shreveport 
Radiology Department 
Proc 13.17 

contrast 
    media/heated). 
*  A larger gauge catheter is 
being used. 
*  Fewer disposables in the 
fluid path (lack  
   of extension tube, stopcock, 
checkvalve,  
   IV extension set).  

Condition #3 

The graph and injection 
start out normal, but 
level off to a less than 
expected peak pressure. 

An inappropriate pressure 
limit may be selected.  If a 
lower pressure is selected 
than what is required to 
deliver the flow rate, the 
injection will start out 
normally for the first few 
seconds.  However, as soon 
as the pressure limit value is 
achieved, the flow rate will 
automatically decrease to 
keep the system from 
exceeding the selected 
pressure limit.   

Check to see if the Pressure 
Graph is approaching the 
dotted horizontal line 
(pressure limit setting).  If it 
is, the injector is pressure 
limiting and causing this 
condition.  See Pressure 
Limiting” in the Operation 
Manual for a thorough 
discussion of causes and 
resolution.   

Note:  Medrad makes no claims about the information on the Pressure Monitor Graph 
other than it is a graphic representation of the force required to move the syringe plunger, 
which closely approximates the pressure developed in the syringe.  

CONDITION# DESCRIPTION

POSSIBLE CAUSES 

POSSIBLE ACTIONS 

Condition #4 

At the start of the 
injection, the 
Pressure Garph 
signal 
immediately 
rises straight up 
and hits the 
pressure limit 
setting (dotted 
horizontal line).  

The fluid path may be 
occluded (after reaching a 
“High Pressure Disarm”. 
Stellant will automatically 
terminate the injection).  
Total occlusions may be 
caused by one of the 
following: 
*  A closed stopcock 
*  An IV extension set with 
the clamp  
   closed off 
*  A defective disposable 
(fluid will not  
   flow through it).   

Consider putting the 
injector on hold and 
investigating.  There is a 
high probability that no 
contrast media is being 
injected into the patient.  

Содержание Stellant D

Страница 1: ...he hand on the ace of the gantry during injection This allows for uninterrupted passage of injected contrast through the axillary and subclavian veins at the thoracic outlet 7 A small volume test inje...

Страница 2: ...wly can result in an error message Piston will automatically advance to the front of the syringe SETUP 1 Using sterile technique attach the sterile loading device and load the syringes by pressing AUT...

Страница 3: ...the remaining volume 4 Armed Indication Lights Contrast green Saline blue will flash when the injector is armed or on hold and will be solid when injecting 5 Initiate the injection at the patient sid...

Страница 4: ...essage cannot be corrected and or the system is not Operating properly DO NOT USE the system Call MEDRAD Service c Routine inspections and maintenance will ensure continued performance and reduce equi...

Страница 5: ...theter is being used An additional disposable is in the fluid path causing a greater restriction to the flow of fluid extension tube stopcock check valve IV extension set Defective syringe under lubri...

Страница 6: ...d causing this condition See Pressure Limiting in the Operation Manual for a thorough discussion of causes and resolution Note Medrad makes no claims about the information on the Pressure Monitor Grap...

Страница 7: ...onnector Tubing extension set or catheter has burst or disconnected from the fluid path fluid is running on the floor The syringe may have failed Fluid is being forced around the syringe plunger inste...

Страница 8: ...nd the LPCT may be connected to the wrong syringe Stop the injector immediately and investigate Check the fluid path for air Condition 8 At the end of a normal injection the pressure graph does not im...

Отзывы: