Q-Sense Technical Reference Manual
Proprietary
37 of 188
4.2
Installation USB Content
The installation USB contains the MEDOC Main Station setup program and additional features
that are required for system normal operation, as detailed below:
1.
MEDOC Main Station setup program.
2.
Software license files.
3.
Technical Reference Manual.
4.
Technical Reference Service Manual.
5.
Remote assistance tools.
The installation USB contains all of the unique system files. Keep it in a
safe place. It may be required for restoring the system configuration in
case of malfunction and/or service during remote control session by
Medoc.
4.3
Setup and Installation Main Steps
The assembly procedure consists of the following steps:
1.
Selecting a location for the system. See section
2.
Connecting the Thermode and any additional accessories to the system. See section
3.
Connecting the laptop and installing the MEDOC Main Station software. See section
page
4.
Configuring the system.
See Q-Sense Technical Reference Service Manual for further details on
the system components.
4.4
Selecting a Location
An appropriate location for the Q-SENSE system should meet the following criteria:
1.
The room temperature should be kept between 18°C and 24°C.
2.
The Q-Sense system should not be placed in a location, which is exposed to direct sunlight
or constant vibrations, or close to heaters or air conditioners.
3.
The space allocated for the system should allow enough room to perform regular system
maintenance procedures.
4.
The room in which the system will be located should be quiet, allowing the tested subject
to concentrate on the examination with minimum distractions.
5.
It is possible to hang the Q-Sense system using Medoc Wall Hanger. To hang the system:
Connect Medoc Wall Hanger to the specific site on the wall.
5.1.
Mark the screw’s holes on the wall, and take the hanger off the wall.
5.2.
Use a 6 mm drill bit to create holes on the wall in the points you marked in b section.
5.3.
Insert masonry anchors to the drilled holes.
5.4.
Connect the hanger by screwing the attached screws to the masonry anchors.
Different type of walls requires different masonry anchors.
Содержание TSA 2001
Страница 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Страница 35: ...Q Sense Technical Reference Manual Proprietary 35 of 188 Figure 7 Q SENSE System Wiring Schema ...
Страница 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Страница 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Страница 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...