Q-Sense Technical Reference Manual
Proprietary
184 of 188
2.1.
Connect the Q-sense system so it would be placed in the MRI control room, while the
Thermode itself is placed in the MRI room, as shown in the diagram:
2.2.
Make sure your PC, Q-sense fMRI unit are all OFF.
2.3.
Install the fMRI thermode Control Room part (see Figure 116) according to the following
instructions:
2.3.1.
Connect the D type panel filter connector on your MRI room penetration panel.
2.3.2.
Connect Q-sense fMRI Connectors to the Q-sense fMRI unit
Figure 116: Q-sense fMRI Thermode Control Room part WA 00281
D type
panel filter
connector
Q-sense Connectors
(To Q-sense unit)
MRI Room
Penetration Panel
Shielded Room
Control Room
D type Panel
filter connector
D type Panel
filter connector
Figure 115- fMRI room setup
Thermode
cable 3m