-34-
Guidance and manufacturer’s declaration - Enclosure Port
The [MDS3001/MDS3001PLUS/MDS3001U] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [MDS3001/MDS3001PLUS/MDS3001U] should assure
that it is used in such an environment.
Phenomenon
Basic EMC
standard
Immunity test levels
Home Healthcare Environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field
IEC 61000-4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
Refer to Proximity fields from RF wireless
communications equipment
Rated power frequency
magnetic fields
IEC 61000-4-8
30A/m
50Hz or 60Hz