Specific warnings

TO REDUCE THE RISK OF ELECTRICAL SHOCK OR INJURY TO PERSONS, DO NOT 

DISASSEMBLE THIS UNIT. Disassembly or attempted repairs if accomplished 

incorrectly can create electrical shock hazard. Only qualified technicians should 

perform servicing of this suction unit.

 

 

Danger

The VAC-ASSIST Suction Unit is a vacuum suction device designed for the collection 

of nonflammable fluid materials in medical applications only. Improper use during 

medical applications can cause injury or death.

For all medical applications:

1. All suctioning should be done in strict accordance with appropriate procedures 

    that have been established by a licensed medical authority.

2.Some attachments or accessories may not fit the tubing supplied. All attachments

    or accessories should be checked prior to use to assure proper fit.

2. AC Power source

 

Caution

The Model HCS7000 is used with a power supply allowing operation on AC voltage 

(115 VAC, 60 Hz). However the correct power source must be used to connect to 

adaptable wall power.

3. Electrical

Compressor:

If the overload protector shuts off the motor frequently, you may have a low 

voltage situation. Low voltage can also be suspected when:

1. The motor does not get up to full power or speed.

2.Other motor operated appliances fail to operate properly. Too many motor 

   operated appliances on same circuit.

 

Note

The motor of this compressor has a thermal overload protector. If the motor 

overheat, the overload protector will shut off the motor. If this occurs, turn the ON/

OFF switch to the OFF position and allow the motor to cool (approximately 5 

minutes). Restart compressor. If the compressor fails to start, wait an additional 5 

minutes before attempting to restart. If this fails, inspect the fuses on your suction 

unit. If the unit still does not restart, call your medical equipment provider 

immediately.
 

Содержание HCS7000

Страница 1: ...VAC ASSIST SUCTION UNIT HCS7000 CAUTION Rx only NOTE Check ALL parts for shipping damage If shipping damage is noted DO NOT use Contact medical equipment provider for further instruction...

Страница 2: ...Introduction 5 5 Illustration of Important Parts 5 6 Set Up 6 7 How to operate your VAC ASSIST Suction Unit 6 8 Cleaning Instructions 9 9 Maintenance 9 10 Troubleshooting 10 11 Disposables and parts...

Страница 3: ...ing 2 Do not place or store product where it can fall or be pulled into a tub or sink 3 Do not place in or drop into water or other liquid 4 Do not reach for a product that has fallen into water Unplu...

Страница 4: ...t 2 AC Power source Caution The Model HCS7000 is used with a power supply allowing operation on AC voltage 115 VAC 60 Hz However the correct power source must be used to connect to adaptable wall powe...

Страница 5: ...duces a maximum vacuum of 560 mmHg The unit is equipped with an 800 ml collection bottle and has been designed for convenient and reliable usages Features New hydrophobic filter design helps contamina...

Страница 6: ...t Please assure that all connections are secure and without leaks before using 7 How to operate your vac assist suction unit 1 Before use check the specification label on the side of the suction unit...

Страница 7: ...to increase vacuum and counter clockwise to decrease vacuum Referring to the vacuum gauge while setting the desired level of vacuum To accurately read the gauge block the patient end of the hose or c...

Страница 8: ...7 Collection canister and sputum should be disposed of per medical waste disposing regulations 8 Install a new collection canister and start to run the suction unit again Note 1 If the unit sustains a...

Страница 9: ...on Operator must wear gloves and medical class mask before disposing of the collection bottle tubing and handle the wasting unit and or accessories 9 Maintenance Suction canister and tubing are dispos...

Страница 10: ...Note Use only VAC ASSIST filters Do not substitute any other material for this hydrophobic filter Improper substitution may lead to contamination or poor performance Fuse Change 1 Use only the same ty...

Страница 11: ...tubing with filter to machine and occlude furthest opening again watching suction level If suction is lost please change filter and tubing and repeat 6 Add canister and repeat suction test If suction...

Страница 12: ...s Collection Bottle Capacity 800 ml cc Operating Temperature 32 104 F 0 40 C Operating Humidity 95 RH Operating Atmospheric Pressure 10 2 15 4 psi 70 106 kpa Storage Temperature 14 158 F 10 70 C Stora...

Страница 13: ...ED WARRANTIES ARE EXCLUDED THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW SOM...

Отзывы: