ACTUATOR REPLACEMENT

HEAD AND FOOT ACTUATORS

1.  Unplug the bed from AC power.
2.  Unplug the actuator cable.
3.  Remove the clevis pin, cotter pin and spacers from both the motor and piston sides of the 

actuator. 

Attention: Note the location of all bushings, washers, cable ties, wire routing and hardware for 
proper reassembly.
4.  Install the new actuator at the motor end first then piston end, using the original clevis pin, 

cotter pin and any spacers that were removed in Step 3.  Installing the motor end first will 
reduce possible damage to spindle and piston mechanism of the actuator.

5.  Plug in the actuator cable. 
6.  Reconnect AC power.
7.  Test the bed by raising and lowering the head and foot sections

HIGH-LOW ACTUATORS

 CAUTION

•  Remove ALL accessories mounted to the bed (assist rails, assist bars and head and 

footboards).  Failure to do so could result in the bed being unstable causing injury.

•  The bed must be turned on its side before replacing actuators.  A minimum of two people 

are required to turn bed on its side.

•  Fully elevate the bed deck and head section before turning the bed on its side to create a 

more stable work platform.

1.  Remove all accessories mounted to the bed (assist rails, assist bars and head and 

footboards).

2.  Elevate the bed to its highest position and raise the head and foot sections.
3.  Unplug the bed from AC power.
4.  With at least TWO people, carefully turn the bed on its side.
5.  Remove the clevis pin, cotter pin and spacers from both the motor and piston sides of the 

actuator. 

Attention: Note the location of all bushings, washers, cable ties, wire routing and hardware for 

proper reassembly.

8.  Install the new actuator at the motor end first then piston end, using the original clevis pin, 

cotter pin and any spacers that were removed in Step 5.  Installing the motor end first will 
reduce possible damage to spindle and piston mechanism of the actuator.

6.  Plug in the actuator cable.
7.  With at least TWO people, carefully return the bed to the upright position.
8.  Reconnect AC power.

9.  Re synchronize the self leveling hi low motors by lowering the bed to its lowest position 

and   holding the elevation down button on the hand pendant an additional 5 seconds after 
the bed has reached its lowest elevation

10. Test the bed by raising and lowering elevation.

NOTE: When replacing head and foot actuators, special care needs t be taken to secure all 
wiring with cable ties as necessary to reduce strain from the actuator cables.  Ensure that 
cable do not get caught on any moving parts.

BED SCHEMATIC

Содержание ALTERRA 1385

Страница 1: ...CE MANUAL ALTERRA 1385 LONG TERM CARE BED FCE1385B This product is not made with natural rubber latex Packaging Handling Instructions see Page 8 Shown with optional side rails mattress head footboards...

Страница 2: ...cific potential entrapment zones in hospital bed systems Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment The entrapment zones represent the relationship of the various bed...

Страница 3: ...IS BED WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS CAUTIONS AND INSTRUCTIONS CONTACT YOUR MEDLINE SALES REP OR MEDLINE CUSTOMER SERVICE BEFORE USIN...

Страница 4: ...r transferring in or out of bed DO NOT ALTER THE PLUG ON THE ELECTRICAL CORD If the plug does not fit the existing outlet a correct outlet should be installed by a licensed electrician NEVER operate t...

Страница 5: ...into the head section H frame using a rubber mallet if necessary INSTALLING OPTIONAL HEAD AND FOOTBOARDS Tools Required Philips Screw Driver Rubber Mallet Materials Required Head and Footboards and H...

Страница 6: ...nt 2 C Clips 2 Locate and install C clips into grooves on the rods of the head section assist rails until they are fully engaged Place C clips on each side of the deck section rod tab as shown below 3...

Страница 7: ...in bodily injury The battery back up is a factory sealed unit DO NOT ATTEMPT to open repair unit Contains sealed lead battery Battery must be recycled REPLACING BATTERY Tools Required Flat Head Screw...

Страница 8: ...place power cord Grinding noise heard from one of the actuators Actuator gear is stripped Actuator gear is stripped replace actuator Replace actuator TROUBLESHOOTING CONTINUED Symptom Possible Cause R...

Страница 9: ...work platform 1 Remove all accessories mounted to the bed assist rails assist bars and head and footboards 2 Elevate the bed to its highest position and raise the head and foot sections 3 Unplug the...

Страница 10: ...Section Bed Deck 17 FCE1385017 1 Wall Guard 18 FCE1385018 1 Head Section H Frame 19 FCE1385019 1 Foot Section H Frame 20 FCE1385020 1 Thigh Section Bed Deck 21 FCE1385021 2 Leg Assembly 22 FCE1385022...

Страница 11: ...y examine all welds for cracks 3 Inspect the deck components for any bending damage or excessive wear 4 Check the ends of the actuators and mounting hardware for any bending damage or excessive wear 5...

Страница 12: ......

Страница 13: ...ir Installation and freight costs for replaced or repaired items remain the responsibility of the Customer This warranty does not cover failure due to negligence accident deliberate abuse nor improper...

Отзывы: