TROUBLE SHOOTING 

 

SYMPTOM  

POSSIBLE CAUSE  

 

ACTION

 

(See steps for isolating problems) 

 

No Power  

Unit not connected to A/C 
power 
Cut/damaged/pinched 
power cord Loose plug for 
pendant 
Cut/damaged/pinched 
pendant cord Actuator 
plug not connected  

 

Check plug to receptacle. 
Check Power Cord for damage. 
Check pendant connection to 
control box. Inspect pendant 
cord for damage. Check 
actuator connection to control 
box. Inspect actuator cord for 
damage.  

Actuators 
operate, but 
the wrong 
movement 
occurs  

Motors plugged into 
wrong ports 
Cut/damaged/pinched 
pendant cord  

 

See picture of control box with 
port locations. Inspect pendant 
cord for damage.  

One or more 
Actuators do 
not function  

Loose (or unplugged) 
connection of actuator 
cord to control box. 
Cut/damaged/pinched 
actuator cord. Defective 
Actuator.  

 

Check actuator plug to control 
box. Inspect Actuator cord for 
damage. Replace Actuator. 

 

STEPS FOR ISOLATING ELECTRICAL PROBLEMS  

Order in which to check for problems is as follows: 
Connections - Pendants - Control Boxes - Actuators  
1.

 

Check all connections  

2.

 

Swap pendant with known working pendant 

3.

 

Swap actuator plugs at known working ports 

4.

 

Check actuators with known working control box  

 

IF USE OF THIS TROUBLE SHOOTING GUIDE DOES NOT 
CORRECT THE PROBLEM, PLEASE CALL MEDLINE INDUSTRIES 
INC. TECHNICAL SUPPORT AT 1-800-MEDLINE.

 

 

OPERATIONAL HINTS:  

1.

 

Be sure that the bed is free and clear of all obstructions for full range of 
motion. Damage to the unit and wall will occur if the bed contacts the wall, 
chair rail, other obstruction while in motion.  

Содержание Alterra 1232

Страница 1: ...1 Medline Industries Inc Medline Alterra 1232 Long Term Care Bed Revised on August 3 2007...

Страница 2: ...Page 3 Safety Summary Page 4 Packaging Handling Page 5 Operation Page 6 Battery Backup Control Box Page 7 Troubleshooting Page 8 Bed Schematic Page 9 Parts List Page 10 Re order Guide Page 11 Maintena...

Страница 3: ...NDERSTAND THE WARNINGS CAUTIONS AND THE INSTRUCTIONS CONTACT YOUR MEDLINE SALES REP OR MEDLINE CUSTOMER SERVICE BEFORE USING THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY RESULT Do not use near explos...

Страница 4: ...e bed NEVER permit more than 1 person in on bed at any time The overall weight capacity of this bed is 450 lbs Body weight should be evenly distributed over the surface of the bed DO NOT lay sit or le...

Страница 5: ...not to cut or nick cords or cables 3 With at least TWO people gently roll bed over to upright position inspect plug and cord for any cuts or damage If none plug in and elevate to a comfortable work he...

Страница 6: ...g of the bed BED OPERATION Bed Positioning Bed functions are operated with a six button or eight button in case of sub acute bed hand held pendant control Button functions on the hand held pendant are...

Страница 7: ...ATTERY BACKUP In the event of the power failure the actuator control box is provided with a battery backup feature that enables bed lowering without AC power Please keep the battery backup plugged int...

Страница 8: ...function Loose or unplugged connection of actuator cord to control box Cut damaged pinched actuator cord Defective Actuator Check actuator plug to control box Inspect Actuator cord for damage Replace...

Страница 9: ...etched or in any way stressed Never wind cords around the bed or other objects to compensate for excess cord length Never allow cords to fall under the casters 3 Do no submerge actuators or controls i...

Страница 10: ...2012 1 Baseboard bumper 13 FCE1232013 2 Head foot section motor 14 FCE1232014 2 Hi Low Motor 15 FCE1232015 4 Mattress retainer 16 FCE1232016 1 Hand pendant 17 FCE1232017 4 Head foot board bracket 18 F...

Страница 11: ...intenance and cleaning procedures should be performed at least once every 3 months This form may be copied and used as a procedural guide and documentation Date______________ Bed Serial ______________...

Страница 12: ...and actuator cords for chafing cuts or excessive wear 3 Make sure all plugs are fully seated and free of damage 4 Make sure all cords are free of moving parts 5 Check all bed functions Head raises an...

Страница 13: ...13 Maintenance Schedule for Alterra 1232 Long Term Care Bed Serial _____________________ Date Inspection Maintenance Cleaning Initials Additional Notes...

Страница 14: ...ducts including modifications without the expressed written consent of Medline Industries Inc shall void this warranty in its entirety NO FURTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITE...

Отзывы: