Revised on August 3, 2007 

Содержание Alterra 1232

Страница 1: ...1 Medline Industries Inc Medline Alterra 1232 Long Term Care Bed Revised on August 3 2007...

Страница 2: ...Page 3 Safety Summary Page 4 Packaging Handling Page 5 Operation Page 6 Battery Backup Control Box Page 7 Troubleshooting Page 8 Bed Schematic Page 9 Parts List Page 10 Re order Guide Page 11 Maintena...

Страница 3: ...NDERSTAND THE WARNINGS CAUTIONS AND THE INSTRUCTIONS CONTACT YOUR MEDLINE SALES REP OR MEDLINE CUSTOMER SERVICE BEFORE USING THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY RESULT Do not use near explos...

Страница 4: ...e bed NEVER permit more than 1 person in on bed at any time The overall weight capacity of this bed is 450 lbs Body weight should be evenly distributed over the surface of the bed DO NOT lay sit or le...

Страница 5: ...not to cut or nick cords or cables 3 With at least TWO people gently roll bed over to upright position inspect plug and cord for any cuts or damage If none plug in and elevate to a comfortable work he...

Страница 6: ...g of the bed BED OPERATION Bed Positioning Bed functions are operated with a six button or eight button in case of sub acute bed hand held pendant control Button functions on the hand held pendant are...

Страница 7: ...ATTERY BACKUP In the event of the power failure the actuator control box is provided with a battery backup feature that enables bed lowering without AC power Please keep the battery backup plugged int...

Страница 8: ...function Loose or unplugged connection of actuator cord to control box Cut damaged pinched actuator cord Defective Actuator Check actuator plug to control box Inspect Actuator cord for damage Replace...

Страница 9: ...etched or in any way stressed Never wind cords around the bed or other objects to compensate for excess cord length Never allow cords to fall under the casters 3 Do no submerge actuators or controls i...

Страница 10: ...2012 1 Baseboard bumper 13 FCE1232013 2 Head foot section motor 14 FCE1232014 2 Hi Low Motor 15 FCE1232015 4 Mattress retainer 16 FCE1232016 1 Hand pendant 17 FCE1232017 4 Head foot board bracket 18 F...

Страница 11: ...intenance and cleaning procedures should be performed at least once every 3 months This form may be copied and used as a procedural guide and documentation Date______________ Bed Serial ______________...

Страница 12: ...and actuator cords for chafing cuts or excessive wear 3 Make sure all plugs are fully seated and free of damage 4 Make sure all cords are free of moving parts 5 Check all bed functions Head raises an...

Страница 13: ...13 Maintenance Schedule for Alterra 1232 Long Term Care Bed Serial _____________________ Date Inspection Maintenance Cleaning Initials Additional Notes...