Steps to take: 
1.  Press Down to switch off the alarm. 
2.  Contact the nearest Medklinn service centre to arrange for maintenance service immediately.

3.  Meanwhile, the device will still continue to operate for the time stated on the Display Screen.

SET

OZONE GENERATOR

FAULT INDICATORS

OXYGEN CHAMBER

AIR FLOW

ALARM BYPASS

Call dealer for service

SET

OZONE GENERATOR

FAULT INDICATORS

OXYGEN CHAMBER

AIR FLOW

ALARM BYPASS

Mode: Temporary
Remain: 30 days

Содержание PRO AS750D

Страница 1: ...PRO AS750D USER MANUAL...

Страница 2: ...Panel Identification 5 Getting Started Display Screen Menu Changing the Settings Fault Indicator on the control panel Controlling the Active Oxygen emission rate Installation of device Installation of...

Страница 3: ...nd Opening the device or tampering with it in any way without proper training will cause the Warranty to become void If you encounter any problem with the device please contact the nearest authorized...

Страница 4: ...y the ventilation holes Doing so will cause excessive heat build up and may result in a fire and irreparable damage to the device 3 Ensure free flow of air around the device 4 Avoid locations where th...

Страница 5: ...r Lock Controls the amount of oxygen flowing into the device s ozone module Filters the air supply Collects any liquid formed from air condensation Outlet for the generated Active Oxygen Controls Moni...

Страница 6: ...wer Supply Fan HVPS Ozone Module Pushes air into the device Increases oxygen purity in the air supply Supplies power to the fan Cools internal compartment temperature Supplies power to the ozone modul...

Страница 7: ...eveling Feet Port for the ozone sensor cable Optional accessory Holes for air to flow in out of the inner compartment Adjustable feet to control height of device on a surface Ozone Sensor Port Levelin...

Страница 8: ...es the air is taken in by the compressor through the air inlet into the HEPA filter This air is forced through a fine mesh filter that traps pollutants before passing into the oxygen tower which compr...

Страница 9: ...To switch the device ON OFF and works as BACK button in the menus SET OZONE GENERATOR FAULT INDICATORS OXYGEN CHAMBER AIR FLOW ALARM BYPASS Display Screen Power Button Power Indicator Up Button Down B...

Страница 10: ...blem with the air flow Alarm Bypass Shows that an alarm bypass was done and the device will continue running for another 30 days before shutting down Fault indicators SET OZONE GENERATOR FAULT INDICAT...

Страница 11: ...device 2 Press the Power Button on the Control Panel to switch on the device 3 The Power Indicator and Display Screen will light up when the device is switched on 4 The device s information status wi...

Страница 12: ...MODE TEMPORARY REMAIN 30 days c MODE STOPPED Shows that the device has stopped operating Make a call to the nearest Medklinn s service center to address the issue MODE STOPPED Call for service 3 P1 Sh...

Страница 13: ...me end time and shows that this current program is being disabled 6 P24h Shows the setting of Program 24h It includes the start time end time and the time interval of the device alternating from on to...

Страница 14: ...sensor 2 10 O3 LVL at ROOM3 Optional Activated when ozone sensor is installed Shows the ozone level detected by ozone sensor 3 O3 LVL at ROOM1 0 00PPM 0 00PPM O3 LVL at ROOM2 0 00PPM O3 LVL at ROOM3 1...

Страница 15: ...the other components stop operating while in this mode i Changing the clock time 4 Press Set on Clock Setting 5 Press Up to change hour and Down to change minutes The time format is in 24 hour clock...

Страница 16: ...gs of Program 24h 10 To set a single operation time for the entire 24 hours press Down until Program 24h is shown and follow steps 12 16 only 9 Press Set to change the Program Setting 11 To set multip...

Страница 17: ...LWAYS Time 24 Disable 10minON 10minOFF 10minON 20minOFF 10minON 30minOFF 10minON 40minOFF 10minON 50minOFF 20minON 20minOFF 20minON 30minOFF 20minON 40minOFF 20minON 50minOFF 30minON 10minOFF 30minON...

Страница 18: ...beside the values to show that this particular time is selected 16 Press Power twice to go back to main menu 10min ON 30min OFF Program 1 Edit Program 1 ON Time Program 1 17 Press Set to configure se...

Страница 19: ...LWAYS Time 24 Disable 10minON 10minOFF 10minON 20minOFF 10minON 30minOFF 10minON 40minOFF 10minON 50minOFF 20minON 20minOFF 20minON 30minOFF 20minON 40minOFF 20minON 50minOFF 30minON 10minOFF 30minON...

Страница 20: ...e Edit Value 08 00 00 22 Press Down and then Set to enter Start Time configuration 23 Press Up to change hour and Down to change minutes Setting Accepted Press any key 24 Press Set to confirm on the s...

Страница 21: ...threshold cut off if applicable 31 Repeat steps 17 30 for Program 2 and or 3 if necessary NOTE If Program 24h is enabled Program 24h will be priority followed by Program 1 Program 2 and Program 3 NOT...

Страница 22: ...tings for Sensor 1 35 Press Up Down to increase decrease parts per million ppm levels by 0 01 The default value is at 0 05 ppm The minimum and maximum values are shown below Min 0 01 ppm Max 1 00 ppm...

Страница 23: ...39 to configure settings on Sensor 2 and Sensor 3 NOTE Only applies to Medklinn PRO AS750D devices which have more than 1 sensor 41 Press Power to go BACK to the previous menu 42 Press Power again to...

Страница 24: ...It is also accompanied with a buzzer sound It directs the user to which component has failed Ozone Generator Fault Condition Buzzer sound emitted red light on Ozone Generator Display Screen shows ALAR...

Страница 25: ...ance service immediately 3 Meanwhile the device will still continue to operate for the time stated on the Display Screen SET OZONE GENERATOR FAULT INDICATORS OXYGEN CHAMBER AIR FLOW ALARM BYPASS Call...

Страница 26: ...mediately Oxygen Chamber Fault Condition Buzzer sound emitted red light on Oxygen Chamber Display Screen shows ALARM O2 Press v Down to mute SET OZONE GENERATOR FAULT INDICATORS OXYGEN CHAMBER AIR FLO...

Страница 27: ...tion Buzzer sound emitted red light on Air Flow Display Screen shows Mode Stopped Call for service device has stopped operating 3 Meanwhile the device will still continue to operate for the time state...

Страница 28: ...is not press Power to switch the device off and contact the nearest Medklinn service centre to arrange for maintenance service immediately Controlling the Active Oxygen emission rate The Medklinn PRO...

Страница 29: ...Default cut off level is at 0 05 parts per million ppm When the Active Oxygen levels reach the value limit the ozone sensor will stop the device from operating The device resumes operation once its l...

Страница 30: ...ch runs through it Channels the Active Oxygen to different Straight Nozzles Channels the Active Oxygen from the device to the Straight Nozzles Outlet for the Active Oxygen to the environment Secures t...

Страница 31: ...Nozzle 3 Lay down the Electrical Conduits along the planned path 4 Slot in the FEP Tube through the Electrical Conduits NOTE Do not bend more than 30 when laying down the FEP Tubes as it will damage t...

Страница 32: ...ured to the ceiling by using the Nylon Butterfly Anchor and two Stainless Steel screws 1 Turn off the AC power outlet Make sure the light at the ON OFF Button is not lit 2 Remove the Fan Filter by pul...

Страница 33: ...pe the Fan gently with a dry cloth 4 Assemble the Fan Filter in proper order and reattach it to the Fan 5 Clean the Fan Filter once a month or when it is clogged up with dirt dust which may block the...

Страница 34: ......

Страница 35: ......

Отзывы: