TECHNICAL NOTE

How to use the RENATRON Reprocessing Connector

The Reprocessing Connector is a barrier connector that isolates a 
dialyzer from the RENATRON

®

 or RENCLEAR

®

 Dialyzer Header Cleaning 

System. A unique quick disconnect feature allows for fast and easy 
connections.
To use the Reprocessing Connector a RENACLEAR  must have the 
modified header cleaner assemblies that attach to the Reprocessing 
Connector. A RENATRON Reprocessing System must first be calibrated  
for use with the Reprocessing Connector. This is accomplished by 
performing a calibration test using the Reprocessing Connector. 
Recalibrate the RENATRON if necessary following the RENATRON 
Directions for Use.

1. 

The Reprocessing Connectors should 
be disinfected with either 1% RENALIN

®

 

solution or full strength ACTRIL

®

 solution  

for 

30 minutes

 

before use.

2. 

The Reprocessing Connector should first be 
attached to the dialyzer blood port connectors.

3. 

The dialyzer can now be attached to the 
RENACLEAR Header Cleaner Assembly or the 
Colder Connectors on the arterial inlet and 
venous outlet lines of a RENATRON.

0

30

15

RENACLEAR Modified 
Header Cleaner 
Assembly

RENATRON Colder 
Connector

Отзывы: