TECHNICAL NOTE
How to use the RENATRON Reprocessing Connector
The Reprocessing Connector is a barrier connector that isolates a
dialyzer from the RENATRON
®
or RENCLEAR
®
Dialyzer Header Cleaning
System. A unique quick disconnect feature allows for fast and easy
connections.
To use the Reprocessing Connector a RENACLEAR must have the
modified header cleaner assemblies that attach to the Reprocessing
Connector. A RENATRON Reprocessing System must first be calibrated
for use with the Reprocessing Connector. This is accomplished by
performing a calibration test using the Reprocessing Connector.
Recalibrate the RENATRON if necessary following the RENATRON
Directions for Use.
1.
The Reprocessing Connectors should
be disinfected with either 1% RENALIN
®
solution or full strength ACTRIL
®
solution
for
30 minutes
before use.
2.
The Reprocessing Connector should first be
attached to the dialyzer blood port connectors.
3.
The dialyzer can now be attached to the
RENACLEAR Header Cleaner Assembly or the
Colder Connectors on the arterial inlet and
venous outlet lines of a RENATRON.
0
30
15
RENACLEAR Modified
Header Cleaner
Assembly
RENATRON Colder
Connector