Without using the transfer pump
• Turn off the water supply to the filtration system.
• Remove and fill each housing with approximately 38 ounces (1.12 liters) of high-level disinfectant and
replace on header (see Filtration Replacement)
• Allow the filters to soak for the time recommended on the product label of the high-level disinfectant for
high-level disinfection.
• Turn on water supply to the filtration system
• Press START and allow the wash cycle to run in order to rinse the disinfectant out of the system. When
all of the water has drained, press STOP.
5