User Manual TDO Sahara 

 
 

5 Operation 

 
 

5.1 Switching the device on  

 

Press the mains switch on the rear of the Instrument. The TDO Sahara is now in standby mode, which 

is indicated with a point in the display. The devices saves the latest switching state when it is turned 

off with the mains switch, so this state will be activated when it is turned on again. 

  

Press the Power Key (Figure 4 (8)) fort wo secounds to activate the TDO Sahara. The device is now in 

default, where it shows the current temperature of the drying chamber.  

 

 

5.2 Temperature Controller Elements and Functions 

 

The temperature controller consists of eight elements. 

 

 

 

 

 

 

 

 

 

 

Page 10                                                                                      

                

Rev. 01 (06/2013)

 

 

Medite GmbH . Wollenweberstr. 12 . 31303 Burgdorf . Germany 

Phone +49 5136 88 84 0 . Fax +49 5136 88 84 55 . [email protected] . www.medite.de 

 

 

(3) 

(1) 

(3)

 Warning light „Hot surface“ 

(4) 

Set-Time key 

(5)

 Set key 

(6)

 Up key 

(7)

 Down key 

(8)

 On/Off key 

(2) 

(2) 

Digital display 

(1) 

Timer key 

(8) 

Figure 4 

 

(7) 

(4) 

(5) 

(6) 

 

 

 

There is always a beep when you press a key. This beep does not indicate a 
successful operation. Pay attention to the time designations for each feature. 

 

 

 

 

 

 

5.2.1 Modes of Operation 

 

The Slide Drying Oven can operate in two different modes.  

The first mode is continuous operation, which means permanent heating. The previously chosen 

temperature will be held constantly. The timer only has a help function with its acoustic signal.  

The second mode relates to the timer. The device starts heating when the timer is activated with the 

key.  

To activate continuous operation, set the timer to 

99:59

Содержание TDO Sahara

Страница 1: ... 31303 Burgdorf Germany Read manual closely before putting device into operation Keep manual accessible with device Contact customer service in case of questions Specifications are subject to change without prior notice User Manual TDO Sahara Slide Drying Oven ...

Страница 2: ...hanges and addendums to date thereto The above mentioned product has been controlled by an authorized test center and meets the following standards and guidelines EU Directive 2006 42 EEC Machinery EU Directive 2004 108 EEC EMV EU Directive 2006 95 EEC Low Voltage DIN EN ISO 9001 VDE 701 702 Burgdorf 19 03 2013 M Ott Managing Director Page 2 Rev 01 06 2013 Medite GmbH Wollenweberstr 12 31303 Burgd...

Страница 3: ...ntended Use 6 1 4 Authorized Operators 6 2 Safety 2 1 Installation 7 2 2 Working with the TDO Sahara 7 3 Components of the TDO Sahara 3 1 Front and Work Area 8 3 2 Technical Data 8 3 3 Scope of Delivery 8 4 Transport and Installation 4 1 Transport and Unpacking 9 4 2 Installation 9 5 Operation 5 1 Switching the device on 10 5 2 Temperature Controller Elements and Functions 5 2 1 Modes of Operation...

Страница 4: ...nweberstr 12 31303 Burgdorf Germany Phone 49 5136 88 84 0 Fax 49 5136 88 84 55 info medite de www medite de 7 Cleaning Maintenance and Recycling 7 1 Cleaning 13 7 2 Maintenance 13 7 3 Recycling 13 8 Care Declaration 9 Table Of Figures 14 10 Maintenance Schedule 15 ...

Страница 5: ...ms of the purchase contract between Medite and the customer The liability of Medite for mistakes concerning details sketches technical drawings etc stated in this manual is excluded within the frameworks of the respective national legal orders This affects especially all financial losses and secondary failures that come about because or in regard of the stated information in this manual Medite is ...

Страница 6: ... or warning are marked with an information symbol Digits in brackets refer to item numbers in figures Digits 1 1 3 Intended Use The TDO Sahara is made for the purpose of embedding tissue specimens in histology and pathology laboratories Processing of food is not permissible The conditions for operation maintenance and service mentioned in this instruction manual have to be strictly observed 1 4 Au...

Страница 7: ...ite spare parts are to be used Any failure to comply with the approved technical regulations for the operation of technical equipment or any non intended use of the instrument will void the manufacturer s liability Use the device only with the allowed power supply Do not use the device in explosion risk areas Safety devices on device and accessories are not to be removed or modified Do not place t...

Страница 8: ... Dimensions Drying Chamber L W H in mm 414 x 273 x 280 Temperature 20 C to 80 C Weight 28 kg Power Supply 115 V 60 Hz 1250 VA 230 V 50 Hz 1250 VA Fuse T6 3A bei 230V 3 3 Scope of Delivery The standard scope of delivery contains one device one power cord one instruction manual Page 8 Rev 01 06 2013 Medite GmbH Wollenweberstr 12 31303 Burgdorf Germany Phone 49 5136 88 84 0 Fax 49 5136 88 84 55 info ...

Страница 9: ...tion Before starting the instrument please make sure that your mains voltage corresponds to the value indicated on the instrument The instrument is supplied ready to use For connection to mains the included power cord for heating instruments must be used It must be connected to a socket with intact ground wire Do not place any inflammable or combustible material near the instrument The minimum dis...

Страница 10: ...edite GmbH Wollenweberstr 12 31303 Burgdorf Germany Phone 49 5136 88 84 0 Fax 49 5136 88 84 55 info medite de www medite de 3 1 3 Warning light Hot surface 4 Set Time key 5 Set key 6 Up key 7 Down key 8 On Off key 2 2 Digital display 1 Timer key 8 Figure 4 7 4 5 6 There is always a beep when you press a key This beep does not indicate a successful operation Pay attention to the time designations f...

Страница 11: ...he value while Up raises it Save the value by pressing Set Time again Then you will be redirected to default where the current time is displayed If you press Up or Down for more than 30 seconds the value cannot be saved To change the time over a wide span let go of the key you are using after ca 25 seconds then press it again To finish setting the timer always press Set Time 5 2 4 Start timer To s...

Страница 12: ...o medite de www medite de 1 3 2 Figure 5 In case of any other defect or if the described procedure is not successful contact the Medite Service Department First pull out the mains plug 6 2 Error messages Indicated Message Cause Solution F1L blinking Sensor error short circuit Call Service Engineer F1H blinking Sensor error broken sensor Call Service Engineer F2L blinking Sensor error short circuit...

Страница 13: ...ance at least once a year To ensure that the drying oven works as environmental friendly as possible we recommend to change the charcoal filter on the back of the device once a month A twelve month package can be ordered by Cat No 12x67 1006 00 7 3 Recycling At the end of operating life of the TDO Sahara please arrange proper recycling Medite is prepared to arrange such proper recycling against pa...

Страница 14: ...es against any unnecessary risk In case of maintenance or repair we kindly ask the customer to make sure that the device and the accessories in question are in a state that bares no health risks for every person that gets in touch with it in the process This includes everyone who prepares the device for dispatch performs the dispatch or takes and unpacks the delivery and also the Medite service st...

Страница 15: ...age 15 Rev 01 06 2013 Medite GmbH Wollenweberstr 12 31303 Burgdorf Germany Phone 49 5136 88 84 0 Fax 49 5136 88 84 55 info medite de www medite de 10 Maintenance Schedule Date Confirmation Customer Confirmation Medite Remarks ...

Страница 16: ...MEDITE GmbH Wollenweberstr 12 31303 Burgdorf Germany Phone 49 5136 88 84 0 Fax 49 5136 88 84 55 info medite de www medite de ...