the  OFF  position  before  storing  the  battery.  CAUTION:  FULLY 
CHARGE BATTERY BEFORE STORAGE AND EVERY 6 MONTHS! 

 

3. CHARGE LEVEL INDICATOR: By pressing the charge level indicator 
button  (3a)  the  charge  level  of  the  battery  can  be  checked.  The 
green LED (3c) will light up for 4 seconds showing the battery charge 
level,  while  battery  is  powering  your  PAP  device.  If  4  charge 
indicator lights (3c) are lit green, then the battery is fully charged. 
Please  refer  to  the  LO  to  HI  charge  level  indicator  image  (3d)  to 
determine the remaining power level. When the battery is charging, 
the charge indicator LEDs (3c) will light up in sequence the charge 
reaches each level.  When the battery is plugged into the charger, 
but  is  already  fully  charged,  then  all  charge  indicator  lights  are  lit 
solid  green.  Warning  Indication:  Once  CPAP  battery  is  fully 
discharged, the orange warning light (3b) will flash, and the sound 
alarm will beep. Discontinue using your PAP device and charge the 
battery as soon as possible.  

4.  USB  PORT:  This  port  can  be  used  to  power  and  charge  most 
electronic hand-held devices.  

5. DC OUT: This port is used to connect a PAP device to the battery. 
First, select the proper output cable for the PAP device being used. 
Select  Output  Cable  (B)  for  ResMed  S9™,  or  Output  Cable  (C)  for 
ResMed AirSense™ 10. Attach one end of the output cable to the 
PAP device and plug the other end into the “DC OUT” port on the 
battery. 

6. LED Flashlight: LED Flashlight provides light.  

7. LED Flashlight Control Button: This button is used to turn the LED 
Flashlight on and off. Press the LED flashlight control button once to 
turn the flashlight on. Press it again to turn the flashlight off. 

 

8. PROTECTIVE ALARM: Alarm will sound and yellow light will flash 
when the remaining power of the battery is less than 5% or in case 
of battery failure. When alarm sounds, press the Mute button once 
to mute the sound. Yellow LED light stops flashing only after 
battery is charged or battery is turned off. Please note: if alarm 

Содержание Pilot-12 plus

Страница 1: ...LIMENTACI N PORT TIL USTED BATTERIE CPAP ALIMENTATION PORTABLE ENGLISH Thank you for purchasing MEDISTROM Pilot 12 Plus Pilot 24 Plus CPAP Battery Portable Power Supply Please read this product user g...

Страница 2: ...2 ENGLISH page 3 ESPA OL p gina 25 FRAN AIS page 48...

Страница 3: ...2 2 1 PACKAGE CONTENTS Pilot 12 Plus 12 2 2 IMPORTANT 13 2 3 INTENDED USE AND COMPATIBILITY Pilot 12 Plus 13 2 4 CONTROL PANEL Pilot 12 Plus 14 2 5 INSTRUCTIONS FOR USE Pilot 12 Plus 16 2 6 PRODUCT SP...

Страница 4: ...B Output cable for ResMed S9 C Output cable for ResMed AirSense 10 D Connector Tip for ResMed S9 Original Power Adapter E Battery sleeve SOLD SEPARATELY F Car charger for Pilot 24 Plus G Co Pilot 24...

Страница 5: ...urer is not responsible for damage done by the battery to any other device not listed above 1 3 INTENDED USE AND COMPATIBILITY Pilot 24 Plus Pilot 24 Plus is equipped with industrial grade genuine rec...

Страница 6: ...inal Power Adapter is NOT included with Pilot 24 Plus It is provided with the ResMed AirSense 10 PAP Device Plug the ResMed AirSense 10 Original Power Adapter into the battery s CHARGE port Ensure tha...

Страница 7: ...ing the battery with the PAP device Remember to turn the switch to OFF position before storing the battery CAUTION FULLY CHARGE BATTERY BEFORE STORAGE AND EVERY 6 MONTHS AFTERWARD 3 CHARGE LEVEL INDIC...

Страница 8: ...ECTIVE ALARM Alarm will sound and yellow LED will flash when the remaining power of the battery is less than 5 or in case of battery failure When alarm sounds press the Mute button once to mute the so...

Страница 9: ...vided with the ResMed S9 PAP Device Plug the Connector Tip D into ResMed S9 Original Power Adapter Plug the other end of the Connector Tip D into the battery s CHARGE port Ensure that the other end of...

Страница 10: ...ON switch 2 on battery control panel C Using with PAP device as a backup power supply Step 1 Connect the battery to the charger see instructions on charging the battery Step 2 Connect the battery to...

Страница 11: ...hter socket and End 2 into the 24 V converter Step 2 NOTE Please use your Output Cable for ResMed AirSense 10 c that is included with your Pilot 24 Plus CPAP Battery for Step 2 Plug in your Output Cab...

Страница 12: ...Two connector tips choose the tip that fits the PAP device being used with the battery E Battery sleeve F Output Cable for Respironics DreamStation G Output Cable for Respironics RemStar System One S6...

Страница 13: ...12 Plus is NOT COMPATIBLE with any PAP devices manufactured by ResMed even if one of the three connector tips D fits the ResMed device Medistrom is NOT RESPONSIBLE for damages caused as a result of s...

Страница 14: ...r outlet Note The battery can be charged regardless of which position the OFF ON switch is toggled If the battery is charging properly the four green charge indicator LEDs 3c will flash and will light...

Страница 15: ...ound alarm will beep Discontinue using your PAP device and charge the battery as soon as possible 4 USB PORT This port can be used to power and charge most electronic hand held devices 5 DC OUT This p...

Страница 16: ...s fully charged B Using with PAP Device as a portable battery Please Note To increase run time please disconnect the heated humidifier from the PAP device When PAP device is connected with the heated...

Страница 17: ...vice as portable battery 2 6 PRODUCT SPECIFICATIONS Pilot 12 Plus Model Name Pilot 12 Plus Capacity 98 Wh Cell Type Industrial grade genuine rechargeable lithium ion cells manufactured by LG Rated Out...

Страница 18: ...charge in 3 4 hours SECTION 3 USING CO PILOT ADD ON UNIT TO INCREASE THE RUN TIME OF PILOT 24 PLUS PILOT 12 PLUS Pilot 24 Plus and Pilot 12 Plus can be used with an add on unit Co Pilot 24 and Co Pil...

Страница 19: ...c carrier to gain approval to take your Pilot 12 Plus Pilot 24 Plus on board the aircraft SECTION 6 WARRANTY PILOT 24 PLUS PILOT 12 PLUS Please contact your CPAP battery dealer for any warranty relate...

Страница 20: ...PILOT 12 PLUS Please read this product user guide carefully before putting the equipment into operation for the first time Observe the safety instructions Do not expose the Pilot 12 Plus Pilot 24 Plu...

Страница 21: ...ve materials are present Do not operate the device if the housing or AC adapter is damaged Immediately disconnect the damaged device from the power by pulling the plug Do not expose the equipment to e...

Страница 22: ...r Codes Yellow and green LEDs alternate flashing Battery has entered over discharge protection to prevent damage to the cells Contact Medistrom technical support Yellow LED flashing rapidly Incorrect...

Страница 23: ...y use the A C adapter that was included with the Pilot 12 PAP device connected to battery is incompatible Ensure that the PAP device you are using is compatible with the Pilot 12 or Pilot 24 you are u...

Страница 24: ...Ensure that the PAP device you are using is compatible with the Pilot 12 or Pilot 24 you are using If errors persist after taking the recommended action contact Medistrom technical support toll free a...

Отзывы: