1

Open the calibrator package (in new box 
of electrodes). Find the three raised bumps 
on the calibrator package. Peel the clear 
cover away from the bumps. 

Important: Failure to calibrate 
properly will cause incorrect results.

2

Hold the calibrator with the LOT 
number facing you.

3

Insert the calibrator into the electrode port. 
Push it in until it stops. The LOT number
shows on the display window with the
time and date.

4

Check that the LOT numbers on all these items match:

Display window 

Electrode calibrator

Electrode instructions for use 

Electrode foil packet

Blood Glucose Calibration Steps

Blood Glucose Monitoring Steps

1

Insert the three black lines at the end 
of the electrode into the electrode port 
until it stops. The monitor turns on 
automatically. The Apply Blood symbol 
and LOT number appear on the 
display screen. 

2

Check that the LOT number on the 
screen matches the batch of electrodes
being used. Touch the blood drop to
the white area at the end of the 
electrode.

3

Continue to touch the blood 
to the end of the electrode 
until the test begins.

4

Accurate results 
in seconds.