Separation Distances Between SI PCF and Mobile RF Equipment
Recommended separation distances between
portable and mobile RF communications equipment and the SI PCF
The SI PCF is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the SI PCF can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the SI PCF as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 KHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.5 GHz
d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures objects and people.
Содержание SI PCF
Страница 1: ...SI PCF User Manual...
Страница 2: ...Ref DOC03148 Issue 1 Friday 06 January 2017 Page 2 of 81 Intentionally blank...
Страница 7: ...Ref DOC03148 Issue 1 Friday 06 January 2017 Page 7 of 81 1 Introduction 1 Introduction User Manual...
Страница 20: ...Introduction SI PCF User Manual Ref DOC03148 Issue 1 Friday 06 January 2017 Page 20 of 81 Intentionally blank...
Страница 21: ...Ref DOC03148 Issue 1 Friday 06 January 2017 Page 21 of 81 2 Installation 2 Installation User Manual...
Страница 36: ...Installation SI PCF User Manual Ref DOC03148 Issue 1 Friday 06 January 2017 Page 36 of 81 Intentionally blank...
Страница 37: ...Ref DOC03148 Issue 1 Friday 06 January 2017 Page 37 of 81 3 Operation 3 Operation User Manual...
Страница 59: ...Ref DOC03148 Issue 1 Friday 06 January 2017 Page 59 of 81 4 Troubleshooting 4 Troubleshooting User Manual...
Страница 65: ...Ref DOC03148 Issue 1 Friday 06 January 2017 Page 65 of 81 5 Technical 5 Technical User Manual...
Страница 71: ...Ref DOC03148 Issue 1 Friday 06 January 2017 Page 71 of 81 6 Appendices 6 Appendices User Manual...
Страница 80: ...Appendices SI PCF User Manual Ref DOC03148 Issue 1 Friday 06 January 2017 Page 80 of 81 Intentionally blank...