Maintenance 

Maintenance schedule 

The stretcher requires regular maintenance. Build your maintenance schedule and follow it. The 

table below represents the minimum maintenance intervals. 

Activity 

Plan 

Cleaning and disinfection 

After every use 

Inspection (see checklist) 

 

By use frequency 1-25x a month 
maintain inspection every 6 months  

 

By use frequency 26-200x a month 
maintain inspection every 3 months 

 

By use frequency over 200x times a 
month maintain inspection every month 

 

Revision  

MEDIROL recommend doing revision regularly. First revision inspection performs after 2 years and 

then regularly ones a year. 

Cleaning 

 

wash all parts with warm water 

 

remove blood stains with brush or solvent 

 

blow the water with compressed air and wipe with soft cloth 

 

surface treat with universal agent  WD-40 and polish 

Disinfection 

 

common disinfectants based on active and hydrogen peroxide can be used to disinfect 

 

the producer recommends to perform test on hidden unused place if used disinfectant is 
inert to the surface of the surface of the product cover 

Warning

 

Long-term use of disinfectants in excessive concentration without neutralization may change color of 

surface, or damage the surface. For cleaning and disinfection of surface do not use the products 

containing organic solvents such as acetone, toluene and gasoline. 

Oiling 

Lubricate all hinged and sliding joints with the universal product WD-40. Disinfectant may cause 
undesirable degreasing of moving joints. Therefore pay attention to correct lubrication. 

 

 

Содержание Sanero M201

Страница 1: ...Monoblock SANERO M201 User Manual ...

Страница 2: ...11 Side folding grab handles 12 Positioning of the rear legs 13 Positioning of the front legs 14 Unlocking of front wheels whirl 15 Drive into ambulance car 16 Drive into ambulance car 17 Basic Accessories 18 Belt system and mattress with pillow 18 Maintenance 19 Maintenance schedule 19 Revision 19 Cleaning 19 Disinfection 19 Oiling 19 Inspectional Note 20 Monoblock 20 Mattress 20 Belt system 20 T...

Страница 3: ...vided as a service to users All other security measures taken by the user should be under the applicable regulations It is recommended arranging training about the correct use of this product in specific situation before use No part of the manuals delivered with this product may not be reproduced copied transmitted stored in a backup system or translated in any form into another language without p...

Страница 4: ...ibed in this manual Improper maintenance may cause injury Use only original accessories and parts for service Improper parts and service may cause injury Do not tamper with the product design Modifications of product may cause injury or product damage COMPATIBILITY WITH FASTENING SYSTEM Integral part of the monoblock SANERO is fastening system ROLFIX fixing the ambulance cot in ambulance car MEDIR...

Страница 5: ...utside of ambulance car Registered in class Non invasive I Statement Meet basic requirements in common use in accordance with user manual is safe effective and appropriate for providing health care Producer accepted precautions by which ensure conformity of all medical means with the technical documentation and the basic requirements of the Government which apply to it To proof the conformity was ...

Страница 6: ...ended for this product or are not compatible with it damage or failure due to accident improper use misuse or negligence In case of any modification of frame or any parts of the product this warranty expires completely Producer proceeds repair in guarantee time by product s repair or exchange for another Producer does not cover the freight charge for defective product or its parts Production label...

Страница 7: ...rsized wheels absorbing shaking from unevenness Remote steer lock system Roll in loading with two step load release mechanism One four point shoulder harness restraint One belt restraints Anatomic mattress Anatomic pillow Technical Specification Length 195 cm Width 56 cm Height 32 66 74 81 88 95 102 cm Weight without mattress and restraints 43 kg Wheels Diameter 20cm Width 5 cm Articulation Backre...

Страница 8: ...8 Technical Parameters ...

Страница 9: ...9 ...

Страница 10: ...10 Operating Instructions Positioning of the footrest ...

Страница 11: ...11 Positioning of the backrest ...

Страница 12: ...12 Side folding grab handles ...

Страница 13: ...t to decrease the action of force at the positioning locker Handle of monoblock s carriage during legs positioning entry and exit from ambulance car always hold from under as presented at the picture to avoid tripping of the carriage from hands During monoblock riding in any case do not manipulate with lever for back legs release With forward riding might happen unwanted folding of front legs ...

Страница 14: ...t legs It is not allowed to brake the trolley during positioning rear swivel wheels have to find their position themselves During positioning the monoblock always uplift to decrease the action of force at the positioning locker ...

Страница 15: ...15 Unlocking of front wheels whirl Set the brake of monoblock s carriage ...

Страница 16: ...ng drive into ambulance car is possible to take advantege to sequent unlock of the locker front and rear legs in two positions of the operating lever During drive into ambulance car monoblock must be always in basic transport position ...

Страница 17: ...der to reduce action of force at locker for positioning During drive out from ambulance car is necessary hold lever of the lock off legs locker during unfold rear legs Especially care about right unfold of front legs which must be safely secured in unfolded position before front carriage wheels ride down from ambulance car s deck ...

Страница 18: ...18 Basic Accessories Belt system and mattress with pillow Mattress with pillow A 1806 Anatomical mattress B 1803 Anatomical pillow Belt system C 1707 Four point belt D 1712 Cross belt E ...

Страница 19: ...ove blood stains with brush or solvent blow the water with compressed air and wipe with soft cloth surface treat with universal agent WD 40 and polish Disinfection common disinfectants based on active and hydrogen peroxide can be used to disinfect the producer recommends to perform test on hidden unused place if used disinfectant is inert to the surface of the surface of the product cover Warning ...

Страница 20: ...nctional in all positions Mechanism for safety folding and unfolding legs by drive into and out from ambulance car is fully functional Control bar of height trolley safety is unbroken Sides folding grab handles are fully functional Backrest and footrest can be fixed in arbitrary position and is fully functional Monoblock can be easily locked into fastener in ambulance car Optional accessories is u...

Страница 21: ...21 Training Note Date Name Type of training ...

Страница 22: ...22 Note from inspection Date Performed maintenance Name ...

Страница 23: ...ustomer Care If you have other questions about the product its operation or maintenance ask your dealer Our dealers have extensive experience with the products MEDIROL They can help you with your individual questions and find the right solution Copyright MEDIROL Ltd All rights reserved MEDIROL logo is a registered trademark of MEDIROL Ltd ...

Отзывы: