+49 8142 448 460
www.medin-medical.com
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinCNO EN
Rev 12 Stand 06.03.2019
8
von
68
Accuracy: ±1 L/min (in the working range) ±2 L/min (outside of the
working range)
Basis of measurement: DIN 1343 at 0°C, 1013.25 mbar and 0% air
humidity (0/1013)
Oxygen concentration:
Setting range: 21% to 100% (in flow working range)
Measurement range: 21% to 100%
Accuracy ± 3% (Vol.)
MLF – 16 or OOM 102 measurement sensor
Pressure measurement:
Measurement range 0 mbar to 15 mbar - pressure graph
Accuracy ±1.3 mbar
Verification: Redundant measurement by two sensors
Push (inspiration support by the
medinCNO
®
)
Setting range
additional flow during the inspiration push:
Min: 0 L/min
Max: 17.5 L/min minus set basic flow
Duration 200 ms to 2 seconds
Accuracy
± 1 L/min (if the total flow =
basic flow + additional
inspiration flow
- is within the flow working range)
± 2 L/min (if the total flow =
basic flow + additional
inspiration flow
- is outside of the flow working range)
Basis of measurement: DIN 1343 at 0°C, 1013.25 mbar and 0% air
humidity (0/1013)
O
2
flush
Duration: 1 minute
Level: +10, +20, +30 % (vol.) above set target oxygen concentration
or 100% oxygen.
Electronic shut-off valve
In the event of an error, interrupts the flow supply to the patient and
opens the tubing system to the atmosphere
Mechanical overpressure valve
Opening pressure 4 kPa (= 40 mbar)
Operating Time
The medinCNO
®
can be used for continuous, long-term operation up
to 4 weeks without a restart in the interim.
IP class
IP20
2.5.3
Overview of the general technical description
Information on these points can be found in the chapters indicated:
•
Conditions for use, transport and storage: see chapter 3.5
•
Characteristics and accuracies of the device: see chapter 2.5.2
•
Information on the installation of the device: see chapter 3
•
Description of the supply gases required: see chapter 3.3
•
Description of the power supply: see chapter 3.2
•
The device is disconnected from the mains by pulling the power supply unit cable of the
medinCNO
®
out of the electrical outlet. Thus when installing the medinCNO
®
, it should be
ensured that this plug is always freely accessible and can be disconnected from the mains