+49 8142 448 460
www.medin-medical.com
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
OP_medinCNO EN
Rev 12 Stand 06.03.2019
20
von
68
-
Minimum reachable push frequency R
insp
in XX / min caused by apnoea
-
Maximum reachable push frequency R
insp
in XX / min caused by triggering synchronised with
respiration
-
Maximum CPAP value during the push in XX.X mbar
-
Set oscillation frequency in XX Hz
-
Set amplitude level of oscillation in level XX
-
Measured amplitude during the oscillation in XX.X mbar
-
Set push length T
insp
in X.X sec
-
Set apnoea time in XX.X sec
Please note:
When reading the trend data, there may be time delays upon receipt at the computer or data system. For this
reason, when reading the data, wait until the last line (
***;2014-04-07;14:22:51;medin-cno1004;record
complete;
) has been issued. Only then have the trend data been fully received.
3.5
Environmental conditions (operation / transport / storage)
The medinCNO
®
must not be used in rooms at risk of explosion, near to flammable substances or close to
locations exposed to splash water.
Warning:
The medinCNO
®
is not suitable for use near magnetic resonance imaging equipment (MR-unsafe) or
electrosurgical equipment.
During operation, the medinCNO
®
should not be used in the direct vicinity of other electronic devices or
stacked. If use as described above is nonetheless necessary, this device and other devices should be observed
to ensure that they are working properly.
Environmental conditions for the medinCNO
®
during operation:
Temperature:
15°C to 35°C
Relative air humidity: Air humidity 20% to 80% (not condensing)
Moisture:
The device is to be kept dry during operation.
Ambient pressure:
700 hPa to 1100 hPa
Oxygen environment: O
2
content <25%
Altitude:
≤ 2000 m above sea level
Cleanliness:
The hygiene regulations valid for the hospital are to be observed. In
addition, the medinCNO
®
is to be operated in standard hospital ambient
conditions.