Operator console
Figure 5-1: medinCNO
®
controls: a) ON/OFF; b) Buttons for entering basic CPAP parameters; c) Dial for use in
changing and confirming settings; d) Trigger button for manual pushes; e) Alarm reset button; f) Button for
mode-specific and other settings
The medinCNO
®
is operated by means of several buttons and a dial. The usual procedure is to press a button
to select the parameter indicated next to that button on the display, then use the dial to select the required
value. This is confirmed by pressing the dial.
The various buttons are divided into several groups:
-
ON/OFF
: (Figure 5-1–a) to switch the medinCNO
®
on and off
-
Basic CPAP settings:
(Figure 5-1 –b) to select and adjust the two basic CPAP parameters of oxygen
concentration (
FiO
2
) and
Flow
. In addition, to set the additional flow for pushes (
F
insp
) and to trigger
an
O
2
Flush
.
-
Push button (PUSH)
: (Figure 5-1 –d) to trigger a push manually and immediately
-
Alarm reset
: (Figure 5-1 –e) to confirm and briefly acknowledge alarms. By pressing this button, the
current alarm sounds are suppressed for 120 seconds and then reactivated. However, if a new alarm
becomes active during this time, this will also be emitted acoustically.
-
Other settings:
(Figure 5-1 –f) to select and alter mode-specific and other settings