background image

 

Quick Start Guide 

(1) Install the battery; 
(2) Put the new battery in correctly before the first use, and then start the body temperature check; 
(3) When used, the front end of the thermometer is aligned to the forehead, with a distance of 5-
10cm. Make sure there is no hair, sweat, cosmetics or hat worn. And then press the measuring switch 
to start the body temperature check; 
(4) When the change of ambient temperature affects the forehead temperature or when the forehead 
perspires,  please  aim  at  the  earlobe  for  measurement. And  make  sure  there  is  no  hair,  sweat, 
cosmetics or hat worn. 
Note: It’s normal that the measured temperature value maybe different, varying from skin color, 
thickness to different parts of human body, as well as large changes in ambient temperature. The 
more exposed the human body is, the more easily and largely the temperature is affected by ambient 
temperature. 

 

Scope of Application 

The  temperature  of  the  measured  object  is  displayed  by  measuring  the  thermal  radiation  of  the 
forehead. 

Composition 

It is composed of shell, infrared temperature sensor, signal receiving processor, key,  Cob board, 
buzzer and LCD. 

Part Names 

 

Содержание XL-F02

Страница 1: ...Instruction Manual Medical Infrared Thermometer Specification and Model XL F02 Please read the manual before use Period of use three years Version A 0 ...

Страница 2: ...ay Icon Definition 7 Contraindications Precautions Warnings and Suggestive Instructions 8 Daily Maintenance Cleaning Maintenance and Maintenance Interval 9 Names and Contents of Toxic and Harmful Substances or Elements in the Product 9 Parts List 10 Explanation of Signs Conformity Abbreviations etc used in Medical Device Labels 10 Troubleshooting 10 Clinical Accuracy 11 Statement 11 Description of...

Страница 3: ...erspires please aim at the earlobe for measurement And make sure there is no hair sweat cosmetics or hat worn Note It s normal that the measured temperature value maybe different varying from skin color thickness to different parts of human body as well as large changes in ambient temperature The more exposed the human body is the more easily and largely the temperature is affected by ambient temp...

Страница 4: ... 42 9 Maximum Permissible Error The maximum allowable error of thermometer is 0 2 within 35 0 42 0 the maximum allowable error is 0 3 beyond 35 0 42 0 Display Resolution 0 1 Measuring Distance 5cm 10cm Operating Environment Temperature 16 35 Relative Humidity 80 Atmospheric Pressure 86kpa 106kpa Transportation and Storage conditions Temperature 20 55 Relative Humidity 93 Atmospheric Pressure 50kPa...

Страница 5: ... each component and maintain the good performance of the thermometer 2 Keep the power supply sufficient 3 After the product is packed it is allowed to use the common means of transportation but it should avoid rain moisture and mechanical collision of extrusion box 2 Unpacking Open the color box which is with single package take out the plastic box with thermometer and then open the plastic box to...

Страница 6: ...l objects such as coins or keys or in the containers that may cause short circuit 7 In case the electrolyte in the battery accidentally enters the eyes please wash the eyes with plenty of water immediately If there is a risk of blindness and other injuries immediately go to the nearest hospital for treatment 8 In case the electrolyte in the battery is accidentally on the skin or clothes please was...

Страница 7: ... close to the stove and do not let it come in contact with water 7 Please do not bump or drop the product and do not use if it is damaged 8 Hair sweat hat or scarf on forehead will affect the accuracy of measurement figure 9 Please ensure the measurement distance is not farther than 5cm 10cm 10 In order to obtain accurate stable and reliable measurement figure when the room temperature changes gre...

Страница 8: ...ting F2 is the fever alarm F3 is the prompt sound switch F4 is the temperature compensation setting F5 is the infrared ray switch Display Icon Definition Display Icon Definition Icon State Description Electricity Indication Display Low power and only for a short time work Display and flash Extremely low power and impossible for normal work Immediate Replacement of the battery is recommended No dis...

Страница 9: ...ke the body temperature check when the temperature is consistent with the ambient temperature 4 The environment around the tested person should be stable and he she can t be tested in places with large air flow such as fan air conditioner outlet etc 5 Do not use the instrument outdoors or in a sunny place 6 Do not hold the front of the thermometer with your hand when measuring 7 The measuring area...

Страница 10: ...not place the thermometer in a place where there is too high or too low temperature damp or direct sunlight dust or corrosive gas 5 If the thermometer is not to be used for a long time more than 3 months please take out the battery for safekeeping 6 It is recommended that user should maintain the thermometer every 3 months Calibration and maintenance The accuracy of this thermometer has been stric...

Страница 11: ...tion Manual Certificate Warranty card 1 Explanation of Signs Conformity Abbreviations etc used in Medical Device Labels Signs 1 Signs on products and packages Reference for use Attention please refer to the attached documents Type B application equipment Keep dry Low voltage prompt Avoid sun exposure Straight up Fragile please handle it with care Safety Critical Parts List Key parts list MCU senso...

Страница 12: ...ter environment the human body is the more easily and greatly the temperature is affected Statement The temperature of the person is attained by measuring the thermal radiation on the forehead of the human body The clinical accuracy or clinical deviation of the thermometer has got the clinical verification and its safety and effectiveness meet the requirements of relevant national laws and regulat...

Страница 13: ...ition to the cables sold by the manufacturer of the machine as spare parts of internal components the use of accessories and cables other than those specified may cause the increase of emission or decrease of anti interference of the medical infrared thermometer Other Guideline and Statement from Manufacturer Electromagnetic Emission Xl F02 medical infrared thermometer is expected to be used in th...

Страница 14: ...ith synthetic material the relative humidity shall be at least 30 Electrical fast transient burst EFT B GB T 17626 4 2kV power supply cord 1kV For input output lines N A N A Electrical Surge GB T 17626 5 1 kV Differential mode voltage 2 kV Common mode voltage N A N A Voltage sag on power input line short time medium time Break and voltage change GB T 17626 11 5 UT for 0 5 cycles on UT 95 sag 40 UT...

Страница 15: ...ided by transmitter manufacturer in watts W d Recommended isolation distance in meters m b The field strength of the fixed RF transmitter is determined by surveying the electromagnetic field C in each frequency range d should be lower than the compliance level Interference may occur near equipment marked with the following symbols Radiofrequency conduction GB T 17626 6 3 V Effective Value 150 kHz ...

Страница 16: ... buyer or user can prevent electromagnetic interference by maintaining the minimum distance between portable and mobile RF communication equipment transmitter and equipment or system as recommended below Rated maximum output power of transmitter W Isolation distance corresponding to different frequencies of transmitter M 150 kHz 80 MHz d 1 2 80 MHz 800 MHz d 1 2 800 MHz 2 5 GHz d 2 3 0 01 N A 0 12...

Страница 17: ...ilure caused by unauthorized folding and refitting of the product 2 Machine failure caused by careless falling during taking or using 3 Failure caused by improper operation without following the operation method in this manual 4 Failure caused by natural disasters such as earthquake fire etc 5 Failure caused by improper maintenance of maintenance shop not authorized by the company Name Contact Cus...

Страница 18: ...unity Humen Town Dongguan City Guangdong Province P R China Product Technical Requirement No Verification for Canton Instrument 20192071228 Medical device registration certificate No Verification for Canton Instrument 20192071228 Production license No Guangdong food and Drug Administration machinery production license 20193511 Product batch No see the machine body Period of use 3 years Release dat...