Set-up Instructions 

1.

 

Remove the TEMP 3

 from the box; leave the black foam inserts inside the cabinet.  The 

cabinet is setup with the foam inserts for Baxter 1000ml fluid bags. 

2.

 

Locate The Documentor

®

 and remove from the shipping box.  Attach The Documentor

®

 to 

the TEMP 3

 following the instructions below. 

 

Place the TEMP 3

 on its back and remove the two indicated screws. 

 

 

 

 

 

Align the holes in the TEMP 3

 Time/Clock Mounting Bracket to the empty screw 

 

holes.  Secure the mounting bracket using the screws removed in step one. 

 

Remove the 3 screws in the back of The Documentor

®

.  Attach The Documentor

®

 to 

 

Time/Clock Mounting Bracket using these 3 screws. 

3.

 

Locate the wall mounting bracket. 

4.

 

Mount the bracket securely in a convenient place – on a wall or cart.  The TEMP 3

 loaded 

with 6 bags of solution will weigh approximately 30 pounds.   When mounting to a wall, care 
should be taken to secure the bracket using suitable bolts for the wall material to support a 
minimum of 90 lbs.  

5.

 

Move the TEMP 3

 to the desired location, check that the power switch is “off” and plug the 

power cord into an approved outlet.  Turn the unit on. 

6.

 

Using the key switch on the front of the cabinet unlock the doors.  Load the labeled bags of 
solution into the cabinet through the top door and close the door.  The door locks will 
automatically lock in approximately 15 seconds.  If the door locks prior to closing the door, 
unlock the door using the key switch and then close the door again. 

Do not remove 
these screws. 

Front of the TEMP 3

 

Remove these two 
screws 

Back Panel 

Содержание TEMP 3 7000 Series

Страница 1: ...5595 A TEMP 3 FLUID WARMER OPERATORS MANUAL 7000 Series Medical Solutions Inc 3901 Centerview Drive Suite L Chantilly VA 20151 800 435 7600 703 834 0039 Fax www warmiv com ...

Страница 2: ...sed to regulate the heating of the fluid bags and the over temperature safety The top display number in red will indicate the temperature of the bag currently being warmed while the bottom display number in green displays the desired set point temperature During the time that the controller is directing the unit to warm a light will illuminate in the upper left corner of the display The Documentor...

Страница 3: ... Ambient to 43 C Factory Preset Temperature 40 C Ambient Operating Conditions Storage Temperature 40 to 80 C 40 to 176 F Humidity Conditions 0 to 90 up to 40 C non condensing Operating Temperature 10 to 55 C 14 to 131 F Atmospheric Pressure during minimum of 16 288 in Hg Transporting Electrical Requirements Power requirements 110 120 VAC 60 HZ Fused 2 each 2 amp Amperage consumed 1 25 amps Heater ...

Страница 4: ...mounting bracket 4 Mount the bracket securely in a convenient place on a wall or cart The TEMP 3 loaded with 6 bags of solution will weigh approximately 30 pounds When mounting to a wall care should be taken to secure the bracket using suitable bolts for the wall material to support a minimum of 90 lbs 5 Move the TEMP 3 to the desired location check that the power switch is off and plug the power ...

Страница 5: ...cabinet 2 Turn the unit on by depressing the power switch The switch will illuminate when energized The controller will perform a self test and then it will display the current temperature of the bottom bag The factory temperature setting is 40 degrees C To change the temperature setting follow the instructions below 3 The controller will display 2 temperatures in its display The top display numbe...

Страница 6: ... button until the number 8 appears in the fourth position When the display reads 8088 Tap the right button 10 When the display flashes all four digits Tap the right button 11 The display flashes P1 date 12 Tap the right button and the small digits on the left representing the year will flash 13 Tap the left button until the desired number appears and then Tap the right button The digits representi...

Страница 7: ...o qualified service personnel Have a qualified person remove all liquid and dry the components Perform any necessary hospital electrical safety checks before returning the unit to service Trouble Shooting 1 If the unit fails to warm do the following a Check to see that the unit is plugged in and the power switch is on b Check the set point temperature setting so that the controller is directing th...

Страница 8: ...w the unit to operate for a minimum of 1 hour before verifying the temperatures After the allotted time compare the temperature on the controller to the temperature reading on the Fluke Thermometer The two temperatures should be within one degree Celsius of each other If the two temperature readings are within one degree Celsius of each other the temperature controller has been verified If they ar...

Страница 9: ...ar and partial side panel to access the electrical compartment Remove 3 screws 2 on the side and 1 on the back Remove 5 screws 1 on the back and 4 on the side Remove 3 screws on the bottom Service panel location and removal ...

Страница 10: ...5595 A 10 Fuse removal instructions To replace the fuses disconnect the power cord from the unit Use a flat screwdriver to unlock the fuse holder from the Power Inlet Replace the fuses as marked ...

Страница 11: ... CONSEQUENTIAL OR OTHER DAMAGES INCLUDING BUT NOT LIMITED TO LOST PROFITS To enable MSI to administer the Warranty properly Buyer shall 1 register the Warranty using the form provided below and 2 promptly notify MSI of any claim hereunder As a condition to the Warranty Buyer shall use and maintain the Equipment according to the provisions of the Operator s service manual IMPORTANT THIS IS YOUR TEM...

Отзывы: