FOR MULTIPLE PATIENT USE
Cleaning
The Space Chamber Plus
®
should be cleaned at least once a month but can be washed more often if
needed.
1. Remove the base by placing your thumb in the hole then pull and twist at the same time. There is no
need to dismantle the mouthpiece.
2. Wash all parts in warm water with a mild liquid detergent.
3.
Do not rinse
as detergent minimises static charge.
4. Shake out excess water and allow to air dry.
Do not rub dry.
5. Firmly replace base once the Space Chamber Plus
®
is dry.
6. Prime the Space Chamber Plus
®
as described overleaf.
The Space Chamber Plus
®
can be cleaned in a dishwasher but must be primed again
as described overleaf.
If it is not clear to you on how to clean your Space Chamber Plus
®
please ask your healthcare professional
for help.
Replace within 12 months.
Autoclaving - Multiple Patient Use
The Space Chamber Plus
®
should be autoclaved between patients.
1. Clean the Space Chamber Plus
®
as per the above cleaning instructions.
2. Disassemble the Space Chamber Plus
®
fully as illustrated to the right.
3. If washing the Space Chamber Plus
®
prior to autoclaving, it must be rinsed free of
all detergent.
4. The Space Chamber Plus
®
has been tested to meet the minimum Australian Standards of 134
o
C
for 3 minutes.
5. Wrap the Space Chamber Plus® in a cloth or two layers of 1-ply polypropylene wrap (Eg: KC600)
to ensure that it does not come into direct contact with the metal parts of the autoclave.
6. Do not place other objects on the Space Chamber Plus
®
.
–
Autoclave at 134 °C for 3 minutes.
–
Recommended cooling time is 30 minutes.
7. Reassemble the Space Chamber Plus
®
as illustrated to the right. Ensure the mouthpiece clicks into
place.
8. Submerge and agitate in warm water mixed with a mild liquid detergent.
9.
Do not rinse
, as the detergent helps to reduce static charge.
–
Shake out excess water.
Do not rub dry.
Allow to drip/air dry.
Note:
Space Chamber Plus
®
can be autoclaved up to 52 times.
MOUTH PIECE
VALVE RING
VALVE
CHAMBER BODY
END CAP