Operating Manual OXI-dive

TM

1

                              

Page 18 of 19

 

 

 

 

APPENDIX B

 

 

‘Handicant’ (Optional) 

   Pin Indexed Bull Nose Adaptor (PIBN) 

 

OPERATING INSTRUCTIONS

 

WARNING:          Oxygen  v i g o r o u s l y   s u p p o r t s   c o m b u s t i o n .          Keep  c l e a r   o f   f l a m m a b l e  

m a t e r i a l s ,  particularly  oil  or  grease.    No  smoking.    Keep  away  from  sources  of 
ignition.   Use  in well-ventilated area.    Cylinder  contains  high  pressure.   Open  valve 
slowly.   Use  only with equipment designed for  oxygen service. Do  not use any of  this 
equipment  for  purposes  other  than  specified.  Store  all  equipment  in  clean  well-
ventilated area.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

1.        Supply Cylinder:

 

Ensure  the  supply  cylinder  is  labelled  'medical  oxygen'  and  is  colour-coded  with  white  shoulders^.  
Check the cylinder valve for any sign of dirt or grease or damage.   If in doubt refer to supplier. 

^Note: oxygen cylinder colours vary in different countries

 

 

2.        Cylinder to be Filled:

 

Do  not  fill  any  other  cylinder  than  that  owned  by  the  user.  Check  the  cylinder  is  not  corroded  nor 
has any dints  or  cuts,  or  any  sign  of  fire  exposure.   Ensure  that  it  is  labelled  "Medical  Oxygen" 
and  is  colour-coded  for  oxygen.  Check  that  10  years  have  not  elapsed  since  last  test  date,  which  is 
stamped on the cylinder shoulder. Check the cylinder valve is clean from oil or grease and has no signs 
of damage. If  any of  these conditions are  not satisfied return the cylinder to  the nearest Medical Gas 
Test Station for inspection. 

 

3.        Filling

 

a.       Fit "Handicant" to the supply cylinder and firmly hand tighten the  white handwheel. Close the 

pressure release valve of the "Handicant". 

b.        Fit small cylinder to "Handicant".  Ensure the pins on the stirrup locate accurately and tighten 

the T-screw. 

c.         Manually support the cylinder throughout the operation. 
d.        Slowly open the valve of the cylinder to be filled. 
e.       Slowly  open  valve  of  supply cylinder.  Filling  will  take  up  to  3  or  4  minutes. Chattering may 

occur and indicates filling is almost complete. Filling is complete when pressure on gauge stops 
rising. 

f.       Close supply cylinder valve. Check pressure on gauge. If the small cylinder is below 7000 kPa 

pressure it  should  be  topped  up  from a  fresh  supply cylinder.   Repeat steps 3a  to  e  to  top  up 
small cylinder. 

g.       Close small cylinder valve, open the pressure release valve of the "Handicant” and disconnect 

the small cylinder from "Handicant". 

h.        Listen near the small cylinder valve outlet to detect any leakage. 
i.         Remove "Handicant" from supply cylinder. 

 

4          After Use

 

Ensure the "Handicant" is kept in a clean storage area. 

Содержание OXI-dive 1

Страница 1: ...ING MANUAL COPYRIGHT RESERVED 2009 No part of this manual may be reproduced in any form without the written permission of Medical Developments International Limited OXI Dive is a trademark of Medical Developments International Limited Australia Revision 4 Jan 2014 ...

Страница 2: ...ation and must not be used as a substitute for such training Only operators who have been trained in the use of oxygen should use the OXI diveTM 1 Medical Developments International Limited makes no claim that the information practices and procedures given in this manual will warrant correct or adequate treatment See Appendix A for brief descriptions of the OXI dive 2 and OXI dive 3 which are alte...

Страница 3: ...Components 6 4 Using the OXI diveTM 1 9 5 Operation of the Closed Circuit Breathing System 10 6 Cleaning and Decontamination 13 7 Storage and Servicing 13 8 Dimensions 13 9 Spare Parts 14 10 References 15 11 Warranty 16 APPENDICES A Other OXI diveTM Models 17 B Handicant Optional 18 C MTV 100 Manually Triggered Ventilator Optional 19 ...

Страница 4: ...supply is a fundamental concern in diving accidents many of which may occur several hours from a medical facility Divers should ensure that there is sufficient oxygen available to allow an injured diver to breathe oxygen from the site of a potential accident until reaching an appropriate medical facility The OXI diveTM 1 is a portable self contained closed circuit resuscitator that provides facili...

Страница 5: ...d with CO2 absorber Universal twin hose single patient use breathing circuit c w 2 L latex free breathing bag and size 5 mask CPR PRO resuscitation mask with oxygen inlet and 1 way valve Single patient use Non rebreathing therapy mask with safety vent reservoir bag and oxygen tubing Single patient use MTV 100 manually triggered ventilator c w 1 2 m white self store oxygen hose and a diameter index...

Страница 6: ...The MTV 100 Manually Triggered Ventilator optional is also fitted to a self seal valve The cylinder contents gauge has a range from 0 to 30 000 kPa and is clearly marked 1 4 1 2 3 4 and FULL The scale at the lower and upper ends of the operating range is coloured red The cylinder valve connects to the pin index yoke of the KDK85 Autovalve Australian Standard AS 2473 Valves for compressed gas cylin...

Страница 7: ...d openings in the grates minimise channelling to increase the efficiency of absorption This can be verified by the even change of colour of the indicator in the soda lime during use The silicone valves are arranged so that during expiration the gas flow is directed to the rebreathing bag through that part of the soda lime in the right side of the absorber On inspiration the gas is drawn through th...

Страница 8: ...8 L min of oxygen for use with an extra casualty if required The therapy tubing is connected to the non rebreathing therapy mask or the CPR PRO Resuscitation Mask 3 Alternatively the self sealed outlet can be used for attachment of the optional MTV 100 Manually Triggered Ventilator Pelican Case The Pelican Case is designed to be completely airtight and waterproof if it is closed properly and the d...

Страница 9: ...on Never expose cylinder and valve to aggressive chemical or industrial environments that may affect the integrity of the cylinder If the cylinder and valve requires cleaning due to contact with flammable or non flammable contaminants refer the cylinder to a certified gas cylinder test station for cleaning and inspection Check the Oxygen Supply Ensure the KDK85 Autovalve fitted to the oxygen cylin...

Страница 10: ...ry again ii IF the breathing bag collapses there is a leak probably due to an inadequate seal of the face mask Reposition the mask and push the Oxygen Bypass cap to refill the breathing bag If necessary increase the flow rate to compensate for leaks iii If the breathing bag compresses readily and refills on release of pressure the casualty is not breathing Ventilate the casualty by compressing the...

Страница 11: ... atmosphere if the reservoir bag empties during inspiration 1 Explain and reassure the patient what is involved during the administration of oxygen 2 If practicable turn an unconscious breathing patient into the left lateral position Take special care if a spinal injury is suspected 3 Turn the oxygen cylinder valve ON anti clockwise 4 Ensure the oxygen tubing is not kinked and attach it firmly to ...

Страница 12: ...e 30 mm scavenging port to direct toxic gas away from the operator 8 Tilt the casualty s head back and support the jaw to achieve an airway 9 It is usually more effective for the rescuer to be positioned behind the casualty s head and to open the airway using the jaw thrust chin lift technique 10 Carefully monitor the casualty s pulse colour and general condition 11 Monitor the oxygen supply and b...

Страница 13: ...quency of testing should be established according to usage but at least every two 2 months An appropriately trained technician should inspect the OXI diveTM 1 oxygen resuscitation unit at yearly intervals The KDK85 Autovalve should be serviced every three 3 years or more often if frequently in use Contact Medical Developments International Limited Ensure the cylinder is currently within the 10 yea...

Страница 14: ...nder C W Integrated Valve ST NDS 149 Bodok seal CY 7580 35 Oxygen cylinder key wheel with chain CI KAB NTH Disposable breathing circuit with size 5 mask AC 7590 17 Non rebreathing Therapy Mask c w reservoir bag AC CPR PRO CPR PRO Resuscitation Mask SS 7595 38 8 L min oxygen handwheel nipple restrictor assembly RS 7515 MTV MTV 100 Manually Triggered Ventilator c w Self Store Hose ...

Страница 15: ...sts and inspections 4 AS 2473 2 2007 Australian Standard Valves for compressed gas cylinders Part 2 Outlet connections threaded and stem inlet threads 5 AS 2473 3 2007 Australian Standard Valves for compressed gas cylinders Part 3 Outlet connections for medical gases including pin indexed yoke connections 6 AS 3840 1 1998 Australian Standard Pressure regulators for use with medical gases Part 1 Pr...

Страница 16: ... expressed or implied MDI neither assume nor authorize any other person to assume liability in connection with the sale This warranty will not apply to any product that has been subject to accident abuse or misuse The warranty is not applicable when unauthorized repairs or modifications have been attempted or when entire units or parts are damaged by accident misuse or improper handling procedures...

Страница 17: ...o an external oxygen supply Optional pin indexed to bull nose adapter PIBN OXI diveTM 3 Compact unit with similar features to the OXI dive 2 but without the small oxygen cylinder Operates from an external oxygen supply The PIBN adapter is included as standard equipment to connect either pin indexed small oxygen cylinders or large threaded oxygen cylinders Incorporates the MTV 100 Manually Triggere...

Страница 18: ...test date which is stamped on the cylinder shoulder Check the cylinder valve is clean from oil or grease and has no signs of damage If any of these conditions are not satisfied return the cylinder to the nearest Medical Gas Test Station for inspection 3 Filling a Fit Handicant to the supply cylinder and firmly hand tighten the white handwheel Close the pressure release valve of the Handicant b Fit...

Страница 19: ... hold the MTV 100 Manually Triggered Ventilator and face mask in place and depress the manual control button until the patient s chest gently rises Then release the button and allow the patient to exhale Repeat this cycle about 12 14 times per minute for an adult and 20 times per minute for a child If the patient begins to breathe spontaneously and triggers the MTV maintain the mask seal and the f...

Отзывы: