Keys, Indicators and Symbols 

 

 

- 8 - 

H. Power Indicator 

The  green  LED  located  on  the  bottom  of  the  oximeter  adjacent  to  the 
POWER CONNECTION illuminates when the Model 5340V is connected 
to electric power. 
 

I. Data Port   

 

 

 

The DATA  PORT

 

is used for serial and analog output. The 

DATA  PORT

 

should  be  used  only  for  connection  to 

equipment that complies with CSA/IEC/UL601-1.  

  To  eliminate  risk  of  shock,  take  care  not  to  touch  the  DATA  PORT  

 

8-pin Connector and the animal simultaneously. 

 

J. Alarm Off Key    

 

 

One  short  depression  of  the  Alarm  Off  Key  silences  the 
alarm  for  a  period  of  60  seconds.  One  long  depression  of 
the  ALARM  OFF  KEY  completely  disables  the  audible 
alarm.  The  LED  displays  show 

A L O F F

,  the  ALARM  OFF 

INDICATOR  begins  to  flash,  and  the  pulse  oximeter  reverts  to  normal 
monitoring.  Silenced alarms can be reactivated by one short depression 
of the ALARM OFF KEY. Disabled audible alarms are reactivated by one 
long depression of the ALARM OFF KEY. The LED displays show 

A L O n

 

K. Alarm Off Indicator 

 

The  ALARM  OFF  INDICATOR  illuminates  and  remains  on  constantly 
when  alarms  are  silenced,  and  flashes  when  audible  alarms  are 
disabled. 
 

L. Oxygen Saturation Alarm Key 

 

 

One  short  depression  of  the  OXYGEN  SATURATION 
ALARM KEY

 

toggles the display between the high and low 

NOTE 

Visual alarm indicators cannot be deactivated. 

 

Содержание 5340V

Страница 1: ...odel 5340V Pulse Oximeter for Veterinary Use User s Manual MEDIAID INC 17517 Fabrica Way Cerritos California 90703 USA Tel 714 367 2848 Fax 714 367 2852 Web www mediaidinc com Email info mediaidinc co...

Страница 2: ......

Страница 3: ...Oximeter for Veterinary Use User s Manual MEDIAID INC 17517 Fabrica Way Cerritos California 90703 USA Tel 714 367 2848 Fax 714 367 2852 Web www mediaidinc com Email info mediaidinc com Mediaid Inc Al...

Страница 4: ...ii...

Страница 5: ...en Saturation Indicator 10 Pulse Rate Indicator 10 Temperature Indicator 10 Alarm Indicator 10 2 3 SYMBOLS 12 Chapter 3 Operating the Model 5340V 3 1 MEASURING PULSE OXIMETRY 13 3 2 MEASURING BODY TEM...

Страница 6: ...2 TROUBLESHOOTING 25 Chapter 5 Equipment Specifications 5 1 OXYGEN SATURATION PULSE RATE TEMPERATURE 27 5 2 ALARMS 27 5 3 ANALOG OUTPUTS 27 5 4 SERIAL COMMUNICATION 27 5 5 GENERAL 28 5 6 SENSORS 28 5...

Страница 7: ...ght detected by a photodiode The pulse oximeter measures within the pulse waveform the ratio of transmitted red to infrared light and thereby determines the oxygen saturation of arterial blood The non...

Страница 8: ...or arterial oxygen saturation heart rate and body temperature The principal features of the Model 5340V are as follows Displays with two bright Light Emitting Diodes LEDS Indicates oxygen saturation h...

Страница 9: ...ring opera tion of the Model 5340V Do not attempt to modify or repair the Model 5340V Dispose of the Model 5340V according to governmental regulations 1 7 Environmental Cautions Do not use in the pres...

Страница 10: ...Principe d Utilisation 4...

Страница 11: ...Keys Indicators and Symbols 5 Chapter 2 Keys Indicators Symbols 2 1 Keys and Indicators...

Страница 12: ...displays the animal s oxygen saturation levels The display is updated with every normal pulse A blinking display signals that SpO2 monitoring may be adversely affected and indicates the continuous pr...

Страница 13: ...ensors with RJ12 connectors are compatible with the Model 5340V To connect a sensor to the oximeter align the plug with the jack on the oximeter and insert it gently until an audible click is heard in...

Страница 14: ...a period of 60 seconds One long depression of the ALARM OFF KEY completely disables the audible alarm The LED displays show ALOFF the ALARM OFF INDICATOR begins to flash and the pulse oximeter reverts...

Страница 15: ...ECREMENT KEY The alarm settings are retained in memory until changed by the user The pulse rate alarm factory settings are high 140 BPM and low 50 BPM The Model 5340V reverts to normal animal monitori...

Страница 16: ...he oxygen saturation level a falling pitch indicates a decreasing oxygen saturation level whereas a raising pitch indicates an increasing oxygen saturation level Audible Pulse Rate Indicator A beeping...

Страница 17: ...st volume MEDIUM PRIORITY alarms are caused by measuring problems such as a disconnected sensor too thin or thick tissue or a faulty sensor The LOW PRIORITY alarm tones sound for one second every 10 s...

Страница 18: ...ats Per Minute BPM Power On Off KEY Alarm Off KEY Oxygen Saturation Alarm KEY Pulse Rate Alarm KEY Increment KEY Decrement KEY Data Port Sensor Cable Connector Power Connection Low Battery Indicator A...

Страница 19: ...s the Internal Confidence Tests see section Internal Confidence Tests in Chapter 3 6 Check that all LED segments light up during the POWER UP TEST 7 Be sure that the VISUAL PULSE RATE INDICATOR is fla...

Страница 20: ...nes to prevent faulty readings and sensor or oximeter complications 5 Power on the unit with one short depression of the POWER ON OFF KEY The Model 5340V tests the displays and indicators with the ini...

Страница 21: ...readings Keep animal movement to a minimum Route sensor cords carefully Avoid applying excessive tension to the sensor or sensor cable 3 4 Setting the Volume Level of the Audible Indicator The Model 5...

Страница 22: ...inactivity 2 Read the high oxygen saturation alarm level with two short depressions of the OXYGEN SATURATION ALARM KEY The red light of the HI ALARM INDICATOR lights up Adjust the high alarm level wit...

Страница 23: ...he ALARM OFF KEY The ALARM OFF INDICATOR stops flashing The LED displays show AL ON 3 7 Power The Model 5340V can be powered by either a power adaptor or the internal rechargeable NiCad battery Batter...

Страница 24: ...ected normal monitoring started Bit 4 Error flag indicating that an error is occurring Bit 3 to 0 Number of bytes in the data packet The SpO2 byte contains the current oxygen saturation level It is in...

Страница 25: ...g data follow these steps 1 Obtain the appropriate DATA PORT Analog Cable from Mediaid Inc 2 Verify that the data receiving device complies with CSA IEC UL 601 1 3 Verify that the Model 5340V is power...

Страница 26: ...release of the POWER ON OFF KEY NOTE Write down the instrument s software version number in the space provided in section Product Information in Chapter 7 This number may enable Mediaid Inc to better...

Страница 27: ...d three volume levels Carefully listen for tone pitch and volume changes Audible Visual Alarm Test Test 3 This sequence tests the instrument s alarm functions The audible alarm tone and the four HI LO...

Страница 28: ...ter is required To perform this test follow these steps 1 Connect one lead to PIN 6 OXYGEN SATURATION ANALOG OUTPUT and the common output lead to PIN 7 GROUND SIGNAL COMMON of the DATA PORT see Chapte...

Страница 29: ...nal circuitry set up values A sensor needs to be applied to a subject The instrument performs normal setup and monitoring functions and periodically displays single digit setup numbers 0 for thin tiss...

Страница 30: ...Principe d Utilisation 24...

Страница 31: ...RR 5 Improper sensor application area of sensor application too opaque extraneous material obstructing the sensor LED and or the photodetector ERR 135 Ambient light too bright direct light onto sensor...

Страница 32: ...Principe d Utilisation 26...

Страница 33: ...2 Digits 69 60 3 Digits Less than 60 unspecified Pulse 32 250 BPM 2 Digits Temperature 0 54 Fahrenheit 0 3 Celsius 5 2 Alarms High SpO2 51 100 Low SpO2 50 99 High Pulse 31 245 BPM Low Pulse 30 244 BP...

Страница 34: ...heric Pressure 1026 377 hPa 770 282 45 mm Hg Relative Humidity 5 to 95 noncondensing Operating Temperature 32 104 Fahrenheit 0 40 Celsius Storage Transport Temperature 22 l49 Fahrenheit 30 65 Celsius...

Страница 35: ...osed with this warranty will conform to the manufacturer s specifications and shall be free from defects in workmanship and materials for a period of five 5 years from the date of original purchase It...

Страница 36: ...n a manner contrary or inconsistent with the User s Manual will void the warranty 6 4 Owner s Registration To assist Mediaid Inc in serving the user please complete the Warranty Registration Card that...

Страница 37: ...2 17517 Fabrica Way Suite H Cerritos CA 90703 USA E mail info mediaidinc com 7 2 Product Information To better assist customers Mediaid Inc recommends all users to write down all pertinent product and...

Страница 38: ...Principe d Utilisation 32...

Страница 39: ...______________________________ Address ____________________________________________________________ ___________________________________________________________________ ________________________________...

Страница 40: ......

Отзывы: